J Chromatogr B Analyt Technol Biomed Life Sci. 2021 Oct 15;1183:122969. doi: 10.1016/j.jchromb.2021.122969. Epub 2021 Oct 4.
Accurate measurement of testosterone is important for the diagnosis of gonadal disorders in men, women, and children. Testosterone measurement has limited accuracy at low concentrations by most commercially available immunoassays. We aimed to develop an LC-MS/MS assay to address the inaccuracy of the in-house immunoassay observed over the past decade and to replace it with the new assay. Testosterone in serum/plasma was extracted with commercial supported liquid extraction plates. Method validation was performed following the CLSI C62-A guideline. A total of 126 samples were used for method comparison between the Beckman UniCel DxI immunoassay and LC-MS/MS. Results by immunoassay were 20% lower compared with LC-MS/MS and had minimal correlation (R = 0.403) with LC-MS/MS below 100 ng/dL. When comparing specimens from the Accuracy-Based Survey from the College of American Pathologists, the newly developed assay agreed well with the CDC reference measurement procedure. In summary, immunoassay measurement of testosterone can be significantly inaccurate, especially at low concentrations. The newly developed LC-MS/MS assay provides accurate results across the entire measurable range.
准确测量睾酮对于男性、女性和儿童性腺疾病的诊断非常重要。大多数市售的免疫分析法在低浓度时对睾酮的测量准确性有限。我们旨在开发一种 LC-MS/MS 测定法,以解决过去十年中观察到的内部免疫分析法的不准确性,并将其替换为新的测定法。血清/血浆中的睾酮用商业支持的液体萃取板进行提取。方法验证遵循 CLSI C62-A 指南进行。共使用 126 个样本对贝克曼 UniCel DxI 免疫分析法和 LC-MS/MS 之间的方法进行比较。与 LC-MS/MS 相比,免疫分析法的结果低 20%,并且在低于 100ng/dL 时与 LC-MS/MS 的相关性最小(R=0.403)。当比较美国病理学家学院准确性研究调查中的标本时,新开发的测定法与 CDC 参考测量程序吻合良好。总之,睾酮的免疫分析法测量可能非常不准确,尤其是在低浓度时。新开发的 LC-MS/MS 测定法在整个可测量范围内提供准确的结果。