Faculty of Chemistry and Chemical Technology, University of Ljubljana, Večna pot 113, 1000 Ljubljana, Slovenia.
Faculty of Marine Sciences, Karadeniz Technical University, Çamburnu, 61530 Trabzon, Turkey.
Molecules. 2021 Oct 12;26(20):6156. doi: 10.3390/molecules26206156.
In this study, comparative analyses were carried out with ion chromatography mass-spectrometry (IC-MS/MS) which has no derivatization step, high-performance liquid chromatography (HPLC) technique, as well as two quantitative and two semi-quantitative immunoassays. The results demonstrated that HPLC and quantitative immunoassay methods were well-correlated with IC-MS/MS in determining histamine in various types of fish products. The best correlation was observed with the HistaSure ELISA Fast Track kit (R = 0.9903). More than half of the values (68%) obtained by two methods were also statistically similar. The results of semi-quantitative test kits also supported histamine values estimated by quantitative methods, with some exceptions. The best results were found for HistaSure Lateral Flow in supporting the quantitative techniques. Therefore, these methods are found suitable for monitoring histamine in fish products in terms of food safety. Good correlations were also observed HPLC and IC-MS/MS in determining cadaverine, putrescine, and tyramine with the highest value observed for tyramine as R = 0.9785. However, no correlation was observed for other biogenic amines, and the majority of the results were significantly different from each other for these amines ( < 0.05). The differences may be caused by the drawbacks reported previously for HPLC. However, further studies are required to confirm the possible effects. This study provides a comparative evaluation of several methods in terms of their suitability in determining biogenic amines in fish products for both monitoring and regulatory purposes.
本研究采用无需衍生化步骤的离子色谱-质谱联用(IC-MS/MS)、高效液相色谱(HPLC)技术以及两种定量和半定量免疫分析法进行了比较分析。结果表明,HPLC 和定量免疫分析法与 IC-MS/MS 测定各种鱼类产品中的组胺具有良好的相关性。与 HistaSure ELISA Fast Track 试剂盒的相关性最好(R = 0.9903)。两种方法中超过一半的(68%)测定值也具有统计学意义上的相似性。半定量检测试剂盒的结果也支持定量方法估计的组胺值,只是存在一些例外。HistaSure 侧向流动法在支持定量技术方面的结果最好。因此,这些方法在食品安全方面被认为适用于监测鱼类产品中的组胺。HPLC 和 IC-MS/MS 测定尸胺、腐胺和酪胺也具有良好的相关性,其中酪胺的相关性最高,R = 0.9785。然而,对于其他生物胺没有观察到相关性,并且这些胺的大多数结果彼此之间存在显著差异(<0.05)。这些差异可能是由 HPLC 先前报道的缺点引起的。然而,需要进一步的研究来确认可能的影响。本研究对几种方法进行了比较评估,以确定它们在监测和监管目的下用于测定鱼类产品中生物胺的适用性。