Xuan Yue, Dvir Danny, Wisneski Andrew D, Wang Zhongjie, Ye Jian, Guccione Julius M, Ge Liang, Tseng Elaine E
Department of Surgery, University of California San Francisco and San Francisco VA Medical Centers, San Francisco, CA.
Division of Cardiology, University of Washington, Seattle, WA.
AsiaIntervention. 2020 Dec;6(2):64-71. doi: 10.4244/aij-d-19-00005.
As indications for transcatheter aortic valve replacement (TAVR) continue to expand towards younger and lower-risk patients, durability becomes an increasingly important question. Durability decreases as leaflet stresses increase, but the impact of transcatheter heart valve (THV) size on stress is unknown. Patient annulus sizes can fall within "grey zones" between 2 TAVR sizes. Our aim was to examine the impact of balloon-expandable THV size on leaflet stresses.
SapienXT 23mm, 26mm, and 29mm sizes (Edwards Lifesciences, Inc) underwent micro-computed tomography scanning to create THV computational models then loaded to systemic pressure using finite element software. THV leaflet maximum principal stresses were 1.69MPa (23mm), 1.70MPa (26mm), and 2.12MPa (29mm) at mean arterial pressure. For intermediate annulus sizes, undersizing the larger THV yielded lower leaflet stresses than oversizing the smaller THV.
Increasing THV size yielded greater leaflet maximum principal stresses, which could suggest a relationship between THV size and long-term durability.. For annulus "grey zones" sizes, undersizing the larger THV resulted in lower leaflet stresses than oversizing the smaller THV. These results may influence optimal device sizing, as THV durability remains an important, unanswered question.
随着经导管主动脉瓣置换术(TAVR)的适应症不断向更年轻、风险更低的患者扩展,耐用性成为一个日益重要的问题。瓣叶应力增加时耐用性会降低,但经导管心脏瓣膜(THV)尺寸对应力的影响尚不清楚。患者瓣环尺寸可能落在两种TAVR尺寸之间的“灰色区域”内。我们的目的是研究球囊扩张式THV尺寸对瓣叶应力的影响。
对23mm、26mm和29mm尺寸的SapienXT瓣膜(爱德华生命科学公司)进行微型计算机断层扫描,以创建THV计算模型,然后使用有限元软件加载至体循环压力。在平均动脉压下,THV瓣叶最大主应力分别为1.69MPa(23mm)、1.70MPa(26mm)和2.12MPa(29mm)。对于中等瓣环尺寸,选择较大THV的小号尺寸产生的瓣叶应力低于选择较小THV的大号尺寸。
THV尺寸增加会导致瓣叶最大主应力增大,这可能表明THV尺寸与长期耐用性之间存在关联。对于瓣环“灰色区域”尺寸,选择较大THV的小号尺寸导致的瓣叶应力低于选择较小THV的大号尺寸。由于THV耐用性仍然是一个重要的、尚未解决的问题,这些结果可能会影响最佳器械尺寸的选择。