Anhui Province Key Laboratory of Advanced Catalytic Materials and Reaction Engineering, School of Chemistry and Chemical Engineering, Hefei University of Technology, Hefei, Anhui, 230009, People's Republic of China.
PLA Army Academy of Artillery and Air Defense, Hefei, Anhui, 230031, People's Republic of China.
Talanta. 2022 Jan 15;237:122954. doi: 10.1016/j.talanta.2021.122954. Epub 2021 Oct 11.
Herein, a simple microfluidic paper-based analytical device (μPAD) by using platinum nanoparticles (Pt NPs) as highly active peroxidase mimic for simultaneous determination of glucose and uric acid was fabricated. The μPAD consisted of one sample transportation layer, four paper-based detection chips, and two layers of hydrophobic polyethylene terephthalate (PET) films. The four detection chips were immobilized with various chromogenic reagents, Pt NPs, and specific oxidase (glucose oxidase or uricase). HO generated by specific enzymatic reactions could oxidize co-immobilized chromogenic reagents to produce colored products by using Pt NPs as efficient catalyst. The multi-layered structure of μPAD could effectively improve the color uniformity and color intensity. Total color intensity from each two detection chips modified with distinct chromogenic reagents were used for quantitative analysis of glucose and uric acid, respectively, resulting in significantly improved sensitivity. The linear range for glucose and uric acid detection was 0.01-5.0 mM and 0.01-2.5 mM, respectively. Satisfied results were obtained for glucose and uric acid detection in real serum samples. An easy-to-use smartphone APP was developed for convenient and intelligent detection. The developed μPAD integrated with smartphone as detector holds great applicability for simple and portable on-site analysis.
本文制备了一种简单的基于金纳米颗粒(Pt NPs)的微流控纸基分析器件(μPAD),用于同时测定葡萄糖和尿酸。μPAD 由一个样品传输层、四个纸基检测芯片和两层疏水的聚对苯二甲酸乙二醇酯(PET)薄膜组成。四个检测芯片分别固定有不同的显色试剂、Pt NPs 和特定的氧化酶(葡萄糖氧化酶或尿酸酶)。特定的酶促反应产生的 HO 可以将共固定的显色试剂氧化,产生有色产物,Pt NPs 作为高效催化剂。μPAD 的多层结构可以有效提高颜色均匀性和强度。使用两种不同显色试剂修饰的两个检测芯片的总颜色强度分别用于定量分析葡萄糖和尿酸,从而显著提高了灵敏度。葡萄糖和尿酸检测的线性范围分别为 0.01-5.0 mM 和 0.01-2.5 mM。在实际血清样本中对葡萄糖和尿酸的检测得到了满意的结果。开发了一个易于使用的智能手机 APP,用于方便和智能的检测。该开发的 μPAD 与智能手机集成作为检测器,具有简单便携的现场分析的巨大适用性。