Umfress Allison C, Glaser Tanya S, Ploysangam Pimpiroon, Freedman Sharon F
Department of Ophthalmology, Duke University Medical Center, Durham, North Carolina.
Department of Ophthalmology, Duke University Medical Center, Durham, North Carolina.
J AAPOS. 2021 Dec;25(6):329.e1-329.e6. doi: 10.1016/j.jaapos.2021.07.010. Epub 2021 Nov 2.
Tonometry guides decision making in children with known or suspected glaucoma. The iCare IC200 is a recently FDA-approved rebound tonometer that can measure intraocular pressure (IOP) in both supine and upright patients. The purpose of this study was to evaluate this new tonometer against more commonly used standard instruments: Tono-Pen in the operating room (OR arm) and Goldmann applanation in the clinic (clinic arm).
The OR arm included sequential children undergoing general anesthesia. The clinic arm included cooperative children seen in clinic. IOP was measured twice (right eye then left eye, repeat) with IC200 and either Tono-Pen (OR arm) or Goldmann (clinic arm).
A total of 99 children were included: 75 for the OR arm and 24 for the clinic arm. There was a strong correlation between the IOP measured by IC200 and each device (R = 0.81 for Tono-Pen; R = 0.78 for Goldmann). In the OR arm, the overall mean IOP difference (ΔIOP [Tono-Pen - IC200]) was -1.6 mm Hg. For eyes with corneal edema (n = 12), ΔIOP ranged from -8 to 15 mm Hg. At IOP >20 mm Hg, mean ΔIOP was -3.7 mm Hg. In the clinic arm, the mean ΔIOP (ΔIOP [Goldmann - IC200]) was -1.2 mm Hg.
In eyes of children under general anesthesia with IOP <20 mm Hg without corneal edema, IC200 and Tono-Pen correlated well. In cooperative clinic patients, Goldmann and IC200 correlated well over the range of IOP studied.
眼压测量指导已知或疑似青光眼患儿的决策制定。iCare IC200是一种最近获得美国食品药品监督管理局(FDA)批准的回弹式眼压计,可测量仰卧位和直立位患者的眼压(IOP)。本研究的目的是将这种新型眼压计与更常用的标准仪器进行比较:手术室中使用的Tono-Pen(手术室组)和诊所中使用的Goldmann压平眼压计(诊所组)。
手术室组纳入接受全身麻醉的连续患儿。诊所组纳入在诊所就诊的合作患儿。使用IC200和Tono-Pen(手术室组)或Goldmann(诊所组)测量眼压两次(先右眼后左眼,重复)。
共纳入99名儿童:手术室组75名,诊所组24名。IC200测量的眼压与每种设备测量的眼压之间存在很强的相关性(与Tono-Pen的R = 0.81;与Goldmann的R = 0.78)。在手术室组中,眼压总体平均差异(ΔIOP [Tono-Pen - IC200])为-1.6 mmHg。对于角膜水肿的眼睛(n = 12),ΔIOP范围为-8至15 mmHg。眼压>20 mmHg时,平均ΔIOP为-3.7 mmHg。在诊所组中,平均ΔIOP(ΔIOP [Goldmann - IC .200])为-1.2 mmHg。
在全身麻醉下眼压<20 mmHg且无角膜水肿的儿童眼中,IC200和Tono-Pen相关性良好。在合作的诊所患者中,在研究的眼压范围内,Goldmann和IC200相关性良好。