在全膝关节置换术中,具有平坦表面的张量会高估中度屈曲松弛度:具有平坦表面的张量和具有插入物形状表面的张量之间的比较。

A tensor with a flat surface overestimates midflexion laxity in total knee arthroplasty: Comparison between a tensor with a flat-shaped surface and a tensor with an insert-shaped surface.

机构信息

Osaka City General Hospital, Osaka City, Osaka, Japan.

Osaka City General Hospital, Osaka City, Osaka, Japan.

出版信息

Knee. 2021 Dec;33:358-364. doi: 10.1016/j.knee.2021.10.018. Epub 2021 Nov 6.

Abstract

BACKGROUND

Soft tissue balance is important for the success of total knee arthroplasty (TKA). Various types of tensors have been developed for the precise measurement of a gap. We hypothesized that the surface shape of the tensor that contacted the TKA component affected the gap measurement. This study aimed to compare the gaps obtained with flat and insert-shaped surface tensors.

METHODS

Two senior surgeons performed 95 TKAs (Vanguard-PS:55 knees; Persona-PS:40 knees). The joint gap was measured in each static knee flexion status (0°, 30°, 45°, 60°, 90°, 120°, and full flexion). We compared the gaps measured with a flat surface tensor and an insert-shape surface tensor. We defined a significant change as a gap difference of >1 mm with a statistical significance.

RESULTS

In Vanguard-PS, significant changes were observed at 30° and 45°. In Persona-PS, significant changes were observed at 30°, 45°, and 60°. In both implants, gaps measured with the flat tensor were larger than those measured with the insert tensor at approximately midflexion, and the significant changes disappeared in higher flexion position over midflexion.

CONCLUSIONS

The surface shape of the tensor affected the measurement of midflexion laxity in TKA. When measuring the gap with a flat tensor, the midflexion laxity was overestimated. A tensor with an insert-shaped surface should be used to measure the gap in TKA.

摘要

背景

软组织平衡对于全膝关节置换术(TKA)的成功至关重要。已经开发出各种类型的张力计来精确测量间隙。我们假设与 TKA 组件接触的张力计的表面形状会影响间隙测量。本研究旨在比较具有平面和插入式表面张力计的间隙。

方法

两名资深外科医生进行了 95 例 TKA(Vanguard-PS:55 膝;Persona-PS:40 膝)。在每个静态膝关节屈曲状态(0°、30°、45°、60°、90°、120°和全屈曲)下测量关节间隙。我们比较了使用平面张力计和插入式表面张力计测量的间隙。我们将>1mm 的间隙差异定义为具有统计学意义的显著变化。

结果

在 Vanguard-PS 中,在 30°和 45°观察到显著变化。在 Persona-PS 中,在 30°、45°和 60°观察到显著变化。在两种植入物中,在大约中屈曲时,使用平面张力计测量的间隙大于使用插入式张力计测量的间隙,并且在中屈曲以上的更高屈曲位置,显著变化消失。

结论

张力计的表面形状会影响 TKA 中中屈曲松弛度的测量。使用平面张力计测量间隙时,中屈曲松弛度会被高估。应使用具有插入式表面的张力计来测量 TKA 中的间隙。