Suppr超能文献

超声闪烁伪像在肾结石检测中的应用:扫描因素对模型的影响

Sonographic twinkle artefact in renal calculus detection: Impact of scanning factors in a phantom model.

作者信息

Haluszkiewicz Elvie, Zelesco Marilyn, Welman Christopher J, Wong Daniel D, Ramsay Duncan

机构信息

Department of Medical Imaging Fiona Stanley Hospital 11 Robin Warren Dr Murdoch Western Australia 6150 Australia.

Department of Diagnostic and Interventional Radiology Royal Perth Hospital Wellington St Perth Western Australia 6847 Australia.

出版信息

Australas J Ultrasound Med. 2017 Aug 23;20(3):106-114. doi: 10.1002/ajum.12053. eCollection 2017 Aug.

Abstract

INTRODUCTION

A common clinical use of the twinkle artefact is to detect and confirm renal calcification on ultrasound. There is however variable demonstration of this artefact using newer generation ultrasound machines. The purpose of this study was to assess the impact of altering multiple scanning factors on ultrasound machines from four major manufacturers on the demonstration of the twinkle artefact. Two custom-made phantoms and five-point Likert-like Scale were utilised.

METHODS

These phantoms contained a range of urinary calculi varying in size, composition, surface contour and depth. The calculi were serially imaged with each ultrasound machine, using manufacturers presets and varying just one imaging factor or parameter at a time. The documented twinkle artefact in the images were subsequently analysed, together with the imaging preset and factor changes that had been made.

RESULTS

Those factors that had the greatest effect in order of impact were colour write priority, colour gain, transducer type, depth of calculus, acoustic power and size of calculus. Variability was also demonstrated between manufacturers.

CONCLUSION

By isolating the effect of scanning factor changes, their importance and contribution to appearances in the ultrasound image can be assessed. Image interpretation in the clinical setting requires an understanding of the underlying physics, particularly in the evaluation of artefacts associated with renal calculi.

摘要

引言

闪烁伪像在临床上的一个常见用途是在超声检查中检测和确认肾钙化。然而,使用新一代超声机器时,这种伪像的显示情况存在差异。本研究的目的是评估改变四个主要制造商生产的超声机器的多个扫描因素对闪烁伪像显示的影响。使用了两个定制的体模和五点李克特量表。

方法

这些体模包含一系列大小、成分、表面轮廓和深度各异的尿路结石。使用各超声机器的制造商预设参数,每次仅改变一个成像因素或参数,对结石进行连续成像。随后分析图像中记录的闪烁伪像以及所做的成像预设和因素变化。

结果

按影响程度排序,对闪烁伪像影响最大的因素依次为彩色写入优先级、彩色增益、换能器类型、结石深度、声功率和结石大小。不同制造商之间也存在差异。

结论

通过分离扫描因素变化的影响,可以评估它们在超声图像外观方面的重要性和贡献。临床环境中的图像解读需要了解基础物理学知识,尤其是在评估与肾结石相关的伪像时。

相似文献

6
Bladder stone: "Must know" ultrasonographic signs.膀胱结石:“必须了解”的超声征象。
Clin Case Rep. 2019 Jan 28;7(3):573-574. doi: 10.1002/ccr3.1977. eCollection 2019 Mar.

本文引用的文献

10
Color Doppler twinkling artifact in hyperechoic regions.高回声区域的彩色多普勒闪烁伪像。
Radiology. 1996 Apr;199(1):269-71. doi: 10.1148/radiology.199.1.8633158.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验