Bas B M, Costongs G M, Janson P C
Department of Haematology, Hospital De Goddelijke Voorzienigheid, Sittard, The Netherlands.
Haemostasis. 1987;17(5):260-7. doi: 10.1159/000215753.
Spectrophotometric clotting assays using chromogenic substrates were adapted for use in a routine laboratory batch analyzer. Contrary to another described method the same plasma dilution can be used for each of the different coagulation parameters. Optimal assay conditions were determined. The within- and between-assay variation, the intraindividual variation during 1 day and from month-to-month were determined; the chromogenic assay was compared to a clotting assay. No significant difference (p less than 0.01) between manual and automated methods was found.