Bas B M, Costongs G M, Janson P C
Department of Haematology, Hospital De Goddelijke Voorzienigheid, Sittard, The Netherlands.
Haemostasis. 1987;17(5):260-7. doi: 10.1159/000215753.
Spectrophotometric clotting assays using chromogenic substrates were adapted for use in a routine laboratory batch analyzer. Contrary to another described method the same plasma dilution can be used for each of the different coagulation parameters. Optimal assay conditions were determined. The within- and between-assay variation, the intraindividual variation during 1 day and from month-to-month were determined; the chromogenic assay was compared to a clotting assay. No significant difference (p less than 0.01) between manual and automated methods was found.
使用发色底物的分光光度法凝血测定法已适用于常规实验室批量分析仪。与另一种描述的方法相反,相同的血浆稀释液可用于每个不同的凝血参数。确定了最佳测定条件。测定了批内和批间变异、一天内和逐月的个体内变异;将发色测定法与凝血测定法进行了比较。未发现手动和自动方法之间存在显著差异(p小于0.01)。