Lysophosphatidylcholine (LPC) plays pivotal roles in several physiological processes and their disturbances are closely associated with various disorders. In this study, we described the development and validation of a reliable and simple flow injection analysis-tandem mass spectrometry (FIA-MS/MS)-based method using dried blood spots (DBS) for quantification of four individual LPC (C20:0, C22:0, C24:0, and C26:0).
Lysophosphatidylcholines were extracted from 3.2 mm DBS with 85% methanol containing 60 ng/ml internal standard using a rapid (30 min) and simple procedure. The analytes and the internal standard were directly measured by triple quadrupole tandem mass spectrometry in multiple reactions monitoring mode via positive electrospray ionization.
Method validation results showed good linearity ranging from 50 to 2000 ng/ml for each LPC. Intra- and inter-day precision and accuracy were within the acceptable limits at four quality control levels. Recovery was from 70.5% to 107.0%, and all analytes in DBS were stable under assay conditions (24 h at room temperature and 72 h in autosampler). The validated method was successfully applied to assessment of C20:0-C26:0LPCs in 1900 Chinese neonates. C26:0-LPC levels in this study were consistent with previously published values.
We propose a simple FIA-MS/MS method for analyzing C20:0-C26:0LPCs in DBS, which can be used for first-tier screening.
溶血磷脂酰胆碱(LPC)在几种生理过程中发挥关键作用,其紊乱与各种疾病密切相关。本研究描述了一种使用干血斑(DBS)定量分析四种个体 LPC(C20:0、C22:0、C24:0 和 C26:0)的可靠且简单的流动注射分析-串联质谱(FIA-MS/MS)方法的开发和验证。
使用含有 60ng/ml 内标物的 85%甲醇从 3.2mm DBS 中提取溶血磷脂酰胆碱,采用快速(30min)简单的程序。通过正电喷雾电离,在三重四极杆串联质谱多重反应监测模式下,直接对分析物和内标物进行检测。
方法验证结果表明,每种 LPC 的线性范围均为 50-2000ng/ml。四个质控水平的日内和日间精密度和准确度均在可接受范围内。回收率为 70.5%-107.0%,在分析条件下(室温下 24 小时和自动进样器中 72 小时),DBS 中的所有分析物均稳定。该验证方法成功应用于 1900 名中国新生儿中 C20:0-C26:0LPC 的评估。本研究中 C26:0-LPC 水平与先前发表的值一致。
我们提出了一种用于分析 DBS 中 C20:0-C26:0LPC 的简单 FIA-MS/MS 方法,可用于一级筛选。