Lasers Surg Med. 2022 Apr;54(4):565-571. doi: 10.1002/lsm.23495. Epub 2021 Nov 22.
To conduct keratometry and investigate the ultrastructural changes after microwave thermokeratoplasty (MTK) in rabbit eyes.
Eighteen New Zealand rabbits (18 eyes) were recruited for this study. Ten eyes were chosen for slit-lamp photography and corneal topography at 1 day, 1 week, 1 month, 3 months, and 6 months postoperatively. Two rabbits were sacrificed on 1 day, 1 week, 1 month, and 3 months postoperatively. All remaining rabbits were sacrificed at 6 months postoperatively. The cornea was observed using hematoxylin and eosin staining and transmission electron microscopy. Data are expressed as mean ± standard deviation. p value was determined using repeated-measures analysis of variance.
Corneal edema, disorganized corneal collagen fibers in the heated area, and necrotic fibroblasts were observed at 1 day and 1 week postoperatively. At 1 month postoperatively, corneal edema was not observed, and corneal cell morphology was normal. Moreover, corneal collagen fibers in the heated area shrunk and tended to be organized. The K1 and K2 values significantly decreased from 49.0 ± 1.2 D and 50.5 ± 0.9 D preoperatively to 40.3 ± 1.2 D and 43.2 ± 0.8 D 6 months postoperatively, respectively. The corneal thickness was 353.1 ± 9.3 µm preoperatively and 317.8 ± 27.7 µm at 6 months postoperatively; the difference was not statistically significant.
Corneal keratometry showed flattening after MTK. Moreover, corneal collagen fibers in the heated area shrunk and tended to be organized at 6 months postoperatively. Further studies are required to determine the safety and stability of MTK.
对兔眼进行角膜曲率计检查和微波热角膜成形术(MTK)后的超微结构变化研究。
本研究纳入 18 只新西兰兔(18 只眼)。10 只眼于术后 1 天、1 周、1 个月、3 个月和 6 个月行裂隙灯照相和角膜地形图检查。术后 1 天、1 周、1 个月和 3 个月处死 2 只兔,术后 6 个月处死其余兔。采用苏木精-伊红染色和透射电镜观察角膜。数据表示为均数±标准差。采用重复测量方差分析比较组间差异。
术后 1 天和 1 周时观察到角膜水肿、加热区角膜胶原纤维排列紊乱和坏死成纤维细胞。术后 1 个月时,角膜水肿消退,角膜细胞形态正常,且加热区的角膜胶原纤维收缩,排列趋于有序。K1 和 K2 值分别由术前的 49.0±1.2 D 和 50.5±0.9 D 显著下降至术后 6 个月的 40.3±1.2 D 和 43.2±0.8 D。术前角膜厚度为 353.1±9.3 μm,术后 6 个月为 317.8±27.7 μm,差异无统计学意义。
MTK 后角膜曲率计显示变平,术后 6 个月加热区的角膜胶原纤维收缩且趋于有序。还需要进一步研究来确定 MTK 的安全性和稳定性。