颅脑创伤患者的颅内压监测与视神经鞘直径的相关性。

Correlation between Invasive Intracranial Pressure Monitoring and Optic Nerve Sheath Diameter in Patients with Traumatic Brain Injury.

机构信息

Department of Surgery -Neurosurgery Unit, Dhulikhel Hospital, Kathmandu University Hospital, Kathmandu University School of Medical Sciences, Dhulikhel, Kavre, Nepal.

Department of Neurosurgery, Upendra Devkota Memorial National Institute, of Neurological and Allied Sciences, Bansbari, Kathmandu, Nepal.

出版信息

Kathmandu Univ Med J (KUMJ). 2021 Apr-Jun;19(74):221-224.

Abstract

Background In management of patients with traumatic brain injury, intracranial pressure holds an important place. Any untoward rise in intracranial pressure portends dreaded complications. Hence, any delay in detecting the issue is considered unacceptable. Objective This study focuses on analyzing the correlation between ultrasound derived optic nerve sheath diameter and intracranial pressure values derived from invasive intracranial pressure monitoring system in a neurosurgical patient with severe traumatic brain injury. Method A prospective observational study was performed using a convenience sample technique including all adult patients with traumatic brain injury who had invasive intracranial monitors placed as part of their clinical care. Ocular ultrasound was performed with 5 - 7.5 MHz linear probe of ACUSON X300 ultrasound system while simultaneous intracranial pressure readings were obtained directly from an invasive intracranial pressure monitoring system. The association between optic nerve sheath diameter and invasive intracranial pressure reading was assessed with the Pearson's correlation coefficient and a receiver operator characteristic curve was created to determine the optimal optic nerve sheath diameter cutoff to detect intracranial pressure > 15 cm H2O. Result One hundred and fifteen ocular ultrasounds were performed on 30 individual patients. The mean age of the population involved in this study is 42.13 years ± 1.89 with male preponderance in the ratio of 6:1. Pearson's correlation coefficient of optic nerve sheath diameter and intracranial pressure was found to be 0.844 (p < 0.000) demonstrating a significant positive correlation. The area under the receiver operating characteristic curve was found to be 0.961 (95% confidence interval = 0.93 to 0.99). Based on analysis of the receiver operating characteristic curve, optic nerve sheath diameter > 4.85 mm performed best to detect intracranial pressure > 15 cm H2 O with a sensitivity of 93.5% and specificity of 83%. Conclusion Optic nerve sheath diameter is a dependable screening tool to evaluate for elevated intracranial pressure among patients with traumatic nerve injury.

摘要

背景

在创伤性脑损伤患者的管理中,颅内压占据着重要地位。任何颅内压的异常升高都预示着可怕的并发症。因此,任何对该问题的检测延迟都被认为是不可接受的。

目的

本研究旨在分析神经外科严重创伤性脑损伤患者中,超声测量的视神经鞘直径与侵入性颅内压监测系统测量的颅内压值之间的相关性。

方法

采用便利抽样技术进行前瞻性观察研究,纳入所有接受侵入性颅内监测作为临床治疗一部分的创伤性脑损伤成年患者。使用 ACUSON X300 超声系统的 5-7.5MHz 线性探头进行眼部超声检查,同时直接从侵入性颅内压监测系统获得颅内压读数。使用 Pearson 相关系数评估视神经鞘直径与侵入性颅内压读数之间的关系,并绘制受试者工作特征曲线,以确定最佳的视神经鞘直径截断值来检测颅内压>15cmH2O。

结果

对 30 名患者进行了 115 次眼部超声检查。本研究人群的平均年龄为 42.13±1.89 岁,男女比例为 6:1。视神经鞘直径与颅内压的 Pearson 相关系数为 0.844(p<0.000),表明两者之间存在显著的正相关关系。受试者工作特征曲线下面积为 0.961(95%置信区间=0.93-0.99)。根据受试者工作特征曲线分析,视神经鞘直径>4.85mm 对检测颅内压>15cmH2O 的效果最佳,其敏感性为 93.5%,特异性为 83%。

结论

视神经鞘直径是评估创伤性脑损伤患者颅内压升高的可靠筛查工具。