Center of Radiotherapy, National Institute of Oncology, Budapest, Hungary.
Institute of Nuclear Techniques, Budapest University of Technology and Economics, Budapest Hungary.
Radiol Oncol. 2021 Nov 19;55(4):508-515. doi: 10.2478/raon-2021-0046.
In the case of dynamic radiotherapy plans, the fractionation schemes can have dosimetric effects. Our goal was to define the effect of the fraction dose on the plan quality and the beam delivery.
Treatment plans were created for 5 early-stage lung cancer patients with different dose schedules. The planned total dose was 60 Gy, fraction dose was 2 Gy, 3 Gy, 5 Gy, 12 Gy and 20 Gy. Additionally renormalized plans were created by changing the prescribed fraction dose after optimization. The dosimetric parameters and the beam delivery parameters were collected to define the plan quality and the complexity of the treatment plans. The accuracy of dose delivery was verified with dose measurements using electronic portal imaging device (EPID).
The plan quality was independent from the used fractionation scheme. The fraction dose could be changed safely after the optimization, the delivery accuracy of the treatment plans with changed prescribed dose was not lower. According to EPID based measurements, the high fraction dose and dose rate caused the saturation of the detector, which lowered the gamma passing rate. The aperture complexity score, the gantry speed and the dose rate changes were not predicting factors for the gamma passing rate values.
The plan quality and the delivery accuracy are independent from the fraction dose, moreover the fraction dose can be changed safely after the dose optimization. The saturation effect of the EPID has to be considered when the action limits of the quality assurance system are defined.
在动态放疗计划中,分割方案可能会产生剂量学效应。我们的目标是确定分次剂量对计划质量和射束传输的影响。
为 5 名早期肺癌患者制定了不同剂量方案的治疗计划。计划总剂量为 60Gy,分次剂量分别为 2Gy、3Gy、5Gy、12Gy 和 20Gy。此外,通过优化后改变规定的分次剂量来重新归一化计划。收集了剂量学参数和射束传输参数,以定义计划质量和治疗计划的复杂性。使用电子射野影像装置(EPID)进行剂量测量来验证剂量传输的准确性。
计划质量与所使用的分割方案无关。在优化后可以安全地改变分次剂量,改变规定剂量的治疗计划的传输准确性不会降低。根据基于 EPID 的测量,高分次剂量和剂量率会导致探测器饱和,从而降低伽马通过率。孔径复杂度评分、转台速度和剂量率变化不是伽马通过率值的预测因素。
计划质量和传输准确性与分次剂量无关,此外,在剂量优化后可以安全地改变分次剂量。在定义质量保证系统的动作限值时,必须考虑 EPID 的饱和效应。