Department of Radiology, Massachusetts General Hospital, Boston, Massachusetts, USA.
Br J Radiol. 2022 Feb 1;95(1130):20211108. doi: 10.1259/bjr.20211108. Epub 2021 Dec 8.
To present the median value and 75th percentile of air kerma at the reference point (K), air kerma-area product (KAP), and fluoroscopic time for a large number of fluoroscopically guided interventional (FGI) procedures.
This retrospective study included the consecutive non-coronary FGI procedures from a Radiology department between May 2016 and October 2018 at a large tertiary-care hospital in the U.S. An in-house developed, semi-automated software, integrated with a dictation system, was used to record patient examination information, including K, KAP and fluoroscopic time. The included patient procedures were categorized into procedure types. A software package R (v. 3.5.1, R Foundation) was used to calculate procedure-specific quartiles of radiation exposure.
Based on analysis of 24,911 FGI cases, median value and 75th percentile are presented for each of K, KAP and fluoroscopic time for 101 procedures that can act as benchmark for comparison for dose optimization studies.
This study provides reference levels ( 50th and 75th percentiles) for a comprehensive list of FGI procedures, reflecting an overall picture of the latest FGI studies for diagnosis, targeted minimally invasive intervention, and therapeutic treatment.
This study provides reference levels (50th and 75th percentiles) for the largest number of fluoroscopically guided interventional procedures reported to date (101 procedures), in terms of air kerma at the reference point, air kerma-area product, and fluoroscopic time, among which these quartiles for ≥50 procedures are presented for the first time.
呈现大量透视引导介入(FGI)程序在参考点(K)处的空气比释动能、空气比释动能-面积乘积(KAP)和透视时间的中位数和第 75 百分位数。
本回顾性研究包括 2016 年 5 月至 2018 年 10 月期间美国一家大型三级保健医院放射科连续进行的非冠状动脉 FGI 程序。一个内部开发的、与听写系统集成的半自动化软件用于记录患者检查信息,包括 K、KAP 和透视时间。所包括的患者程序被分类为程序类型。使用 R(v.3.5.1,R 基金会)软件包计算特定于程序的辐射暴露四分位数。
基于对 24911 例 FGI 病例的分析,为 101 种可作为剂量优化研究比较基准的程序,提供了 K、KAP 和透视时间的中位数和第 75 百分位数。
本研究为广泛的 FGI 程序提供了参考水平(第 50 百分位数和第 75 百分位数),反映了最新 FGI 研究在诊断、靶向微创介入和治疗治疗方面的整体情况。
本研究提供了迄今为止报告的最大数量(101 种)透视引导介入程序的参考水平(第 50 百分位数和第 75 百分位数),包括参考点处的空气比释动能、空气比释动能-面积乘积和透视时间,其中首次呈现了 50 多个程序的这些四分位数。