BMC Cancer. 2021 Dec 2;21(1):1288. doi: 10.1186/s12885-021-09029-3.
To evaluate the effect of the single energy metal artifact reduction (SEMAR) algorithm with a multidetector CT (MDCT) for knee tumor prostheses.
First, a phantom of knee tumor prosthesis underwent a MDCT scan. The raw data was reconstructed by iterative reconstruction (IR) alone and IR plus SEMAR. The mean value of the CT number and the image noise were measured around the prosthesis at the stem level and articular level. Second, 95 consecutive patients with knee tumor prostheses underwent MDCT scans. The raw data were also reconstructed by the two methods. Periprosthetic structures were selected at the similar two levels. Four radiologists visually graded the image quality on a scale from 0 to 5. Additionally, the readers also assessed the presence of prosthetic complication and tumor recurrence on a same scale.
In the phantom, when the SEMAR was used, the CT numbers were closer to normal value and the noise of images using soft and sharper kernel were respectively reduced by up to 77.1% and 43.4% at the stem level, and by up to 82.2% and 64.5% at the articular level. The subjective scores increased 1 ~ 3 points and 1 ~ 4 points at the two levels, respectively. Prosthetic complications and tumor recurrence were diagnosed in 66 patients. And the SEMAR increased the diagnostic confidence of prosthetic complications and tumor recurrence (4 ~ 5 vs. 1 ~ 1.5).
The SEMAR algorithm can significantly reduce the metal artifacts and increase diagnostic confidence of prosthetic complications and tumor recurrence in patients with knee tumor prostheses.
评估单能金属伪影减少(SEMAR)算法联合多排 CT(MDCT)对膝关节肿瘤假体的效果。
首先,膝关节肿瘤假体的体模进行 MDCT 扫描。原始数据分别通过迭代重建(IR)和 IR 联合 SEMAR 进行重建。在假体的柄部水平和关节水平处测量假体周围的 CT 值和图像噪声的平均值。其次,95 例膝关节肿瘤假体患者连续进行 MDCT 扫描。原始数据也通过两种方法进行重建。在类似的两个水平选择假体周围结构。四位放射科医生对图像质量进行 0 至 5 分的评分。此外,读者还对假体并发症和肿瘤复发的存在进行相同的评分。
在体模中,使用 SEMAR 时,CT 值更接近正常值,在柄部水平,使用软和锐利核时的图像噪声分别降低了 77.1%和 43.4%,在关节水平分别降低了 82.2%和 64.5%。主观评分分别提高了 1 至 3 分和 1 至 4 分。在两个水平,有 66 例患者诊断出假体并发症和肿瘤复发。SEMAR 提高了假体并发症和肿瘤复发的诊断信心(4 至 5 分与 1 至 1.5 分)。
SEMAR 算法可以显著减少金属伪影,增加膝关节肿瘤假体患者假体并发症和肿瘤复发的诊断信心。