Andersson Karin M, Norrman Eva, Geijer Håkan, Krauss Wolfgang, Cao Yang, Jendeberg Johan, Geijer Mats, Lidén Mats, Thunberg Per
1 Department of Medical Physics, Faculty of Medicine and Health, Örebro University, Örebro, Sweden.
2 School of Health and Medical Sciences, Örebro University, Örebro, Sweden.
Br J Radiol. 2016 Jul;89(1063):20150993. doi: 10.1259/bjr.20150993. Epub 2016 Apr 28.
To evaluate metal artefact reduction (MAR) techniques from four CT vendors in hip prosthesis imaging.
Bilateral hip prosthesis phantom images, obtained by using MAR algorithms for single-energy CT data or dual-energy CT (DECT) data and by monoenergetic reconstructions of DECT data, were visually graded by five radiologists using 10 image quality criteria. Comparisons between the MAR images and a reference image were performed for each scanner separately. Ordinal probit regression analysis was used.
The MAR algorithms in general improved the image quality based on the majority of the criteria (up to between 8/10 and 10/10) with a statistical improvement in overall image quality (p < 0.001). However, degradation of image quality, such as new artefacts, was seen in some cases. A few monoenergetic reconstruction series improved the image quality (p < 0.004) for one of the DECT scanners, but it was only improved for some of the criteria (up to 5/10). Monoenergetic reconstructions resulted in worse image quality for the majority of the criteria (up to 7/10) for the other DECT scanner.
The MAR algorithms improved the image quality of the hip prosthesis CT images. However, since additional artefacts and degradation of image quality were seen in some cases, all algorithms should be carefully evaluated for every clinical situation. Monoenergetic reconstructions were in general concluded to be insufficient for reducing metal artefacts.
Qualitative evaluation of the usefulness of several MAR techniques from different vendors in CT imaging of hip prosthesis.
评估四家CT供应商的金属伪影减少(MAR)技术在髋关节假体成像中的效果。
使用单能量CT数据或双能量CT(DECT)数据的MAR算法以及DECT数据的单能量重建获得双侧髋关节假体模体图像,由五名放射科医生使用10项图像质量标准进行视觉分级。对每台扫描仪的MAR图像与参考图像分别进行比较。采用有序概率回归分析。
基于大多数标准(高达8/10至10/10),MAR算法总体上提高了图像质量,整体图像质量有统计学改善(p < 0.001)。然而,在某些情况下可见图像质量下降,如新伪影。对于其中一台DECT扫描仪,一些单能量重建系列提高了图像质量(p < 0.004),但仅部分标准有所改善(高达5/10)。对于另一台DECT扫描仪,大多数标准下(高达7/10)单能量重建导致图像质量变差。
MAR算法提高了髋关节假体CT图像的质量。然而,由于在某些情况下可见额外伪影和图像质量下降,应针对每种临床情况仔细评估所有算法。总体而言,单能量重建不足以减少金属伪影。
对不同供应商的几种MAR技术在髋关节假体CT成像中的实用性进行定性评估。