Suppr超能文献

手持式超声设备与传统推车式超声机测量腹主动脉的比较。

Abdominal aorta measurements by a handheld ultrasound device compared with a conventional cart-based ultrasound machine.

机构信息

From the Department of Radiology and Medical Imaging, College of Applied Medical Sciences, Prince Sattam bin Abdulaziz University, Alkharj, Saudi Arabia.

From the Department of Rediodiagnostics and Medical Imaging, Prince Sultan Military Medical City, Riyadh, Saudi Arabia.

出版信息

Ann Saudi Med. 2021 Nov-Dec;41(6):376-382. doi: 10.5144/0256-4947.2021.376. Epub 2021 Dec 2.

Abstract

BACKGROUND

Ultraportable or pocket handheld ultrasound devices (HUD) may be useful for large-scale abdominal aortic aneurysm screening. However, the reproducibility of measurements has not been compared with conventional cart-based ultrasound machines.

OBJECTIVES

Investigate the intra- and inter-operator reproducibility of a HUD compared with a conventional ultrasound machine for aortic screening.

DESIGN

Analytical, cross-sectional.

SETTING

Ultrasound department at a large tertiary care hospital in Riyadh.

PATIENTS AND METHODS

Eligible male participants aged ≥60 years were invited to participate upon arriving for a non-vascular ultrasound appointment. Three repeated anteroposterior measurements of the transverse aorta were made at the proximal and distal locations for each machine before repeating the measurements on a subset of participants by a second blinded operator. Intraclass correlation coefficients (ICC) and the Bland-Altman method were used to analyze reproducibility.

MAIN OUTCOME MEASURE

Inter-system and intra- and inter-operator ICCs.

SAMPLE SIZE

114 males with repeated measurements by second operator on a subset of 35 participants.

RESULTS

The median age (interquartile range) of participants was 68 years (62-74 years). The intra- and inter-operator ICCs were all >0.800 showing almost perfect agreement except for the inter-operator reproducibility at the proximal location using a conventional machine (ICC= 0.583, =.007) and the Butterfly device (ICC=0.467, =.037). The inter-system ICCs (95% CI) were 0.818 (0.736-0.874) and 0.879 (0.799-0.924) at the proximal and distal locations, respectively. The mean difference in aortic measurement between the ultrasound systems was 0.3 mm (1.7%) in the proximal location and 0.6 mm (3.6%) in the distal location. In total, >91% of the difference in measurements between the machines was <3 mm. The mean scanning time was 4:16 minutes for the conventional system and 3:53 minutes for the HUD (=.34).

CONCLUSIONS

Abdominal aortic screening using a HUD was feasible and reliable compared with a conventional ultrasound machine. A pocket HUD should be considered for large-scale screening.

LIMITATIONS

No cases of abdominal aortic aneurysm in the sample and lack of blinding.

CONFLICT OF INTEREST

None.

摘要

背景

超便携或口袋式手持式超声设备(HUD)可能有助于大规模的腹主动脉瘤筛查。然而,其测量的可重复性尚未与传统的基于推车的超声机进行比较。

目的

研究 HUD 与传统超声机在主动脉筛查方面的内-操作者和间-操作者的可重复性。

设计

分析性、横断面研究。

地点

利雅得一家大型三级护理医院的超声科。

患者和方法

邀请年龄≥60 岁的合格男性参与者在进行非血管超声预约时参加。在近端和远端位置,使用两种机器对每位参与者进行 3 次重复的横主动脉前后测量,然后由第二位盲法操作者对部分参与者重复测量。使用组内相关系数(ICC)和 Bland-Altman 法分析可重复性。

主要观察指标

系统间、内-和间-操作者 ICC。

样本量

对 35 名参与者中的一部分进行了第二次操作者重复测量,共 114 名男性。

结果

参与者的中位年龄(四分位距)为 68 岁(62-74 岁)。内-和间-操作者 ICC 均>0.800,显示出几乎完美的一致性,只有在近端位置使用传统机器(ICC=0.583,P=0.007)和蝴蝶设备(ICC=0.467,P=0.037)时的操作者间可重复性除外。近端和远端的系统间 ICC 分别为 0.818(0.736-0.874)和 0.879(0.799-0.924)。两种超声系统测量主动脉的平均差值分别为近端 0.3 毫米(1.7%)和远端 0.6 毫米(3.6%)。总体而言,机器间测量差异的>91%<3 毫米。传统系统的平均扫描时间为 4:16 分钟,HUD 为 3:53 分钟(P=0.34)。

结论

与传统超声机相比,使用 HUD 进行腹部主动脉筛查是可行且可靠的。口袋式 HUD 应考虑用于大规模筛查。

局限性

样本中无腹主动脉瘤病例,缺乏盲法。

利益冲突

无。