The aim of this study was to compare the failure rate of two new generation pulse oximeters at different probe positions, and with and without vasoconstriction, in anaesthetised cats.
This prospective clinical study included 103 cats in which the new generation pulse oximeters, the Rad-5 (Masimo) and EDAN H100N (EDAN), were evaluated. Premedication consisted of the vasoconstrictive drug combination butorphanol (0.2 mg/kg IV) and dexmedetomidine (5 µg/kg IV), or butorphanol only (0.2 mg/kg IV). Pulse oximeter failure rate at the tongue was compared between both groups. Pulse oximeter failure rate was also analysed at the alternative probe positions of the lip, pinna, knee fold and toe in the butorphanol group. Student's -test, Wilcoxon matched pairs signed rank test, Mann-Whitney U-test, Friedman test and χ test were performed. A value <0.05 was considered to be statistically significant.
Overall failure to achieve an adequate signal was 37.6% with the Masimo and 48.0% with the EDAN pulse oximeter ( <0.0001). At the standard probe position on the tongue, the Masimo failed in 4.5%, while the EDAN failed in 35.3% ( <0.0001). Vasoactive premedication increased the failure rate for the Masimo from 3.8% to 5.2% ( = 0.3414) and for the EDAN from 22.4% to 49.0% ( <0.0001). At the alternative probe positions of the lip and knee fold, failure rates for the Masimo were lower (39.7% and 81.4%) than with the EDAN (52.6% and 94.4%; = 0.0231 and = 0.0005, respectively), while the Masimo failed more often at the pinna (63.5%) than the EDAN (47.4%; = 0.0044). At the alternative probe position of the toe, the failure rate for the Masimo (32.7%) was not different from the EDAN (38.5%; = 0.7547).
The Masimo pulse oximeter had lower signal failure rates at the standard probe position on the tongue and at 2/4 alternative probe positions. The standard probe position on the tongue had the lowest failure rate for both devices. Dexmedetomidine-induced vasoconstriction increased the failure rate for the EDAN but not for the Masimo pulse oximeter.
本研究旨在比较两种新型脉搏血氧仪在麻醉猫不同探头位置、有无血管收缩时的失败率。
本前瞻性临床研究纳入了 103 只猫,评估了新型脉搏血氧仪 Rad-5(Masimo)和 EDAN H100N(EDAN)。预给药包括血管收缩药物组合布托啡诺(0.2mg/kg IV)和右美托咪定(5μg/kg IV)或仅布托啡诺(0.2mg/kg IV)。比较两组猫舌部的两种脉搏血氧仪的失败率。在布托啡诺组,还分析了唇部、耳郭、膝褶和脚趾等替代探头位置的脉搏血氧仪失败率。采用 Student's t 检验、Wilcoxon 配对符号秩检验、Mann-Whitney U 检验、Friedman 检验和 χ 检验。 值<0.05 被认为具有统计学意义。
总体而言,Masimo 未能获得足够信号的失败率为 37.6%,EDAN 为 48.0%(<0.0001)。在标准探头位置(舌部),Masimo 失败率为 4.5%,而 EDAN 为 35.3%(<0.0001)。血管活性预给药使 Masimo 的失败率从 3.8%增加到 5.2%(=0.3414),EDAN 从 22.4%增加到 49.0%(<0.0001)。在唇部和膝褶等替代探头位置,Masimo 的失败率(39.7%和 81.4%)低于 EDAN(52.6%和 94.4%;=0.0231 和=0.0005),而 Masimo 在耳郭的失败率(63.5%)高于 EDAN(47.4%;=0.0044)。在替代探头位置(脚趾),Masimo 的失败率(32.7%)与 EDAN(38.5%)无差异(=0.7547)。