Magn Reson Med. 2022 May;87(5):2178-2193. doi: 10.1002/mrm.29120. Epub 2021 Dec 14.
Implement a fast, motion-robust pulse sequence that acquires T -weighted, T -weighted, T -weighted, T fluid-attenuated inversion recovery, and DWI data in one run with only one prescription and one prescan.
A software framework was developed that configures and runs several sequences in one main sequence. Based on that framework, the NeuroMix sequence was implemented, containing motion robust single-shot sequences using EPI and fast spin echo (FSE) readouts (without EPI distortions). Optional multi-shot sequences that provide better contrast, higher resolution, or isotropic resolution could also be run within the NeuroMix sequence. An optimized acquisition order was implemented that minimizes times where no data is acquired.
NeuroMix is customizable and takes between 1:20 and 4 min for a full brain scan. A comparison with the predecessor EPIMix revealed significant improvements for T -weighted and T fluid-attenuated inversion recovery, while taking only 8 s longer for a similar configuration. The optional contrasts were less motion robust but offered a significant increase in quality, detail, and contrast. Initial clinical scans on 1 pediatric and 1 adult patient showed encouraging image quality.
The single-shot FSE readouts for T -weighted and T fluid-attenuated inversion recovery and the optional multishot FSE and 3D-EPI contrasts significantly increased diagnostic value compared with EPIMix, allowing NeuroMix to be considered as a standalone brain MRI application.
在一次扫描中仅使用一个方案和一个预扫描,实现一种快速、运动稳健的脉冲序列,以采集 T 加权、T 加权、T 加权、T 液体衰减反转恢复和 DWI 数据。
开发了一个软件框架,该框架可在一个主序列中配置和运行多个序列。基于该框架,实现了 NeuroMix 序列,其中包含使用 EPI 和快速自旋回波(FSE)读出的运动稳健单次激发序列(无 EPI 扭曲)。也可以在 NeuroMix 序列中运行可选的多激发序列,以提供更好的对比度、更高的分辨率或各向同性分辨率。实施了优化的采集顺序,最大限度地减少了没有采集数据的时间。
NeuroMix 具有可定制性,完整的脑部扫描时间在 1:20 到 4 分钟之间。与前身 EPIMix 的比较显示,T 加权和 T 液体衰减反转恢复的性能有显著提高,而在类似配置下仅需多 8 秒的时间。可选对比度的运动稳健性较差,但在质量、细节和对比度方面有显著提高。对 1 名儿科和 1 名成年患者的初步临床扫描显示出令人鼓舞的图像质量。
与 EPIMix 相比,T 加权和 T 液体衰减反转恢复的单次激发 FSE 读出以及可选的多激发 FSE 和 3D-EPI 对比度显著提高了诊断价值,使得 NeuroMix 可被视为一种独立的脑部 MRI 应用。