数字血压计与汞柱血压计在中等收入国家相关性的评估:社会经济状况的作用。

Evaluation of correlation between digital vs. mercury sphygmomanometer in a middle-income country: The role of socio-economic situation.

机构信息

Hypertension Research Center, Cardiovascular Research Institute, Isfahan University of Medical Sciences, Isfahan, Iran.

Isfahan Cardiovascular Research Center, Cardiovascular Research Institute, Isfahan University of Medical Sciences, Isfahan, Iran.

出版信息

Clin Exp Hypertens. 2022 Feb 17;44(2):113-118. doi: 10.1080/10641963.2021.1999047. Epub 2021 Dec 20.

Abstract

BACKGROUND

Using manometer sphygmomanometers as standard measurement tool, there are controversial data regarding accuracy and validity of digital manometers for measurement of systolic (SBP) and diastolic blood pressure (DBP). Thus, we aimed to compare the accuracy of readings of digital sphygmomanometer in reference to mercury sphygmomanometer in a large population of healthy adults.

METHODS AND MATERIALS

This cross-sectional study was performed on 1119 healthy adults. We measured participant's blood pressure once with mercury sphygmomanometer, as gold standard and again with digital mercury sphygmomanometer. Blood pressure was measured in sitting position after 5 min of rest and preferentially from right arm unless deformed.

RESULTS

The mean ± standard deviation of age of participants was 37.25 ± 10.45 years. Majority of participants were male 588 (52.5%). The right/left SBP measured by digital sphygmomanometer were significantly higher compared with those measured by mercury sphygmomanometer: 115.37 ± 12.33 vs 110.95 ± 10.06/113.69 ± 11.77 vs 110.23 ± 10.34, respectively ( < .001), while an opposite result was observed about right/left DBP: 68.60 ± 8.55 vs 70.60 ± 8.31/69.39 ± 8.31 vs 70.75 ± 8.41, respectively ( < .001). In subgroup analysis in terms of marital status, education, and income, we observed similar findings.

CONCLUSION

According to the results of our data analysis, it was shown that the digital device measurements had significant incompatibility with the mercury sphygmomanometers and it seems that digital devices still cannot be used as the gold standard in blood pressure measurement.

摘要

背景

使用压力计血压计作为标准测量工具,关于数字血压计测量收缩压(SBP)和舒张压(DBP)的准确性和有效性存在争议数据。因此,我们旨在比较大量健康成年人中数字血压计读数与水银血压计的准确性。

方法和材料

这是一项横断面研究,共纳入了 1119 名健康成年人。我们使用水银血压计作为金标准,对参与者的血压进行一次测量,然后再次使用数字水银血压计进行测量。测量前,参与者先在休息 5 分钟后取坐位,除非手臂变形,否则优先测量右臂血压。

结果

参与者的年龄平均值±标准差为 37.25±10.45 岁。大多数参与者为男性 588 名(52.5%)。与水银血压计相比,数字血压计测量的收缩压:115.37±12.33 比 110.95±10.06/113.69±11.77 比 110.23±10.34 更高(<0.001),而舒张压的结果则相反:68.60±8.55 比 70.60±8.31/69.39±8.31 比 70.75±8.41 更高(<0.001)。在根据婚姻状况、教育程度和收入进行的亚组分析中,我们观察到了类似的结果。

结论

根据我们数据分析的结果,数字设备的测量与水银血压计存在显著不兼容,似乎数字设备仍不能作为血压测量的金标准。