Utz J A, Herfkens R J, Johnson C D, Shimakawa A, Pelc N, Glover G, Johnson G A, Spritzer C E
AJR Am J Roentgenol. 1987 Mar;148(3):629-33. doi: 10.2214/ajr.148.3.629.
MR images can be obtained with a 2-sec scan time when an extremely short repetition rate (22 msec), limited flip angle (30 degrees), and gradient refocused echoes are used. Comparison of 415 such images obtained in 29 patients with routine T1-weighted (TR 500, TE 25) and T2-weighted (TR 2000, TE 80) images showed that images free of respiratory artifacts could be obtained in all patients. Although abdominal organs were well seen with 2-sec scan time, overall evaluation of these organs was better on routine T1-weighted images. Vascular structures, however, were seen as well or better on the 2-sec images in 60% of cases. The images were extremely sensitive to field nonhomogeneity, and metallic artifact was exaggerated in five patients with surgical clips. Two-sec MR images provide a rapid method of localizing abdominal organs for further evaluation. The sensitivity to blood flow may assist in the assessment of vascular patency.
当使用极短的重复率(22毫秒)、有限的翻转角(30度)和梯度重聚焦回波时,可在2秒的扫描时间内获得磁共振成像(MR)图像。对29例患者获得的415幅此类图像与常规T1加权(TR 500,TE 25)和T2加权(TR 2000,TE 80)图像进行比较,结果显示所有患者均可获得无呼吸伪影的图像。尽管在2秒扫描时间下腹部器官显示良好,但在常规T1加权图像上对这些器官的整体评估更好。然而,在60%的病例中,血管结构在2秒图像上显示得同样好或更好。这些图像对磁场不均匀性极其敏感,并且在5例有手术夹的患者中金属伪影被放大。2秒MR图像提供了一种快速定位腹部器官以便进一步评估的方法。对血流的敏感性可能有助于评估血管通畅情况。