Suppr超能文献

使用 GE Discovery MI PET 前端架构对假设的长轴向视野 PET 扫描仪进行蒙特卡罗评估。

Monte Carlo evaluation of hypothetical long axial field-of-view PET scanner using GE Discovery MI PET front-end architecture.

机构信息

Department of Radiology, University of Iowa Hospitals and Clinics, Iowa City, Iowa, USA.

Department of Physics and Astronomy, University of Iowa, Iowa City, Iowa, USA.

出版信息

Med Phys. 2022 Feb;49(2):1139-1152. doi: 10.1002/mp.15422. Epub 2022 Jan 10.

Abstract

PURPOSE

The development of total-body PET scanners is of growing interest in the PET community. Investigation into the imaging properties of a hypothetical extended axial field-of-view (AFOV) GE Healthcare SiPM-based Discovery MI (DMI) system architecture has not yet been performed. In this work, we assessed its potential as a whole-body scanner using Monte Carlo simulations. The aim of this work was to (1) develop and validate a Monte Carlo model of a four-ring scanner and (2) extend its AFOV up to 2 m to evaluate performance gain through NEMA-based evaluation.

METHODS

The DMI four-ring geometry and its pulse digitization scheme were modeled within the GATE Monte Carlo platform using published literature. The GATE scanner model was validated by comparing results against published NEMA performance measurements. Following the validation of the four-ring model, the model was extended to simulate 8-, 20-, 30-, and 40-ring systems. Spatial resolution, sensitivity, NECR, and scatter fraction were characterized with modified NEMA NU-2 2018 standards; however, the image quality measurements were not acquired due to computational limitations. Spatial resolutions were simulated for all scanner ring configurations using point sources to examine the effects of parallax errors. NEMA count rates were estimated using a standard 70 cm scatter phantom and an extended version of scatter phantom of length 200 cm with (1-800) MBq of F for all scanners. Sensitivity was evaluated using NEMA methods with a 70 cm standard and a 200 cm long line source.

RESULTS

The average FWHM of the radial/tangential/axial spatial resolution reconstructed with filtered back-projection at 1 and 10 cm from the scanner center were 3.94/4.10/4.41 mm and 5.29/4.89/5.90 mm for the four-ring scanner. Sensitivity was determined to be 14.86 cps/kBq at the center of the FOV for the four-ring scanner using a 70 cm line source. Sensitivity enhancement up to 21-fold and 60-fold were observed for 1 and 2 m AFOV scanners compared to four-ring scanner using a 200 cm long line source. Spatial resolution simulations in a 2 m AFOV scanner suggest a maximum degradation of ∼23.8% in the axial resolution compared to the four-ring scanner. However, the transverse resolution was found to be relatively constant when increasing the axial acceptance angle up to ±70°. The peak NECR was 212.92 kcps at 22.70 kBq/ml with a scatter fraction of 38.9% for a four-ring scanner with a 70 cm scatter phantom. Comparison of peak NECR using the 200 cm long scatter phantom relative to the four-ring scanner resulted in a NECR gain of 15 for the 20-ring and 28 for the 40-ring geometry. Spatial resolution, sensitivity, and scatter fraction showed an agreement within ∼7% compared with published measured values.

CONCLUSIONS

The four-ring DMI scanner simulation was successfully validated against published NEMA measurements. Sensitivity and NECR performance of extended 1 and 2 m AFOV scanners based upon the DMI architecture were subsequently simulated. Increases in sensitivity and count-rate performance are consistent with prior simulation studies utilizing extensions of the Siemens mCT architecture and published NEMA measurements with the uEXPLORER system.

摘要

目的

全身体 PET 扫描仪的发展越来越受到 PET 社区的关注。尚未对假设的扩展轴向视野(AFOV)GE Healthcare SiPM 基于 Discovery MI(DMI)系统架构的成像特性进行研究。在这项工作中,我们使用蒙特卡罗模拟评估了其作为全身扫描仪的潜力。这项工作的目的是:(1)开发和验证四环扫描仪的蒙特卡罗模型;(2)将其 AFOV 扩展到 2m 以通过 NEMA 评估来评估性能增益。

方法

在 GATE 蒙特卡罗平台中,使用发表的文献对 DMI 四环几何形状及其脉冲数字化方案进行建模。通过将结果与发表的 NEMA 性能测量值进行比较,验证了 GATE 扫描仪模型的准确性。在验证了四环模型之后,将模型扩展到模拟 8、20、30 和 40 环系统。使用修改后的 NEMA NU-2 2018 标准对空间分辨率、灵敏度、NECR 和散射分数进行了表征;然而,由于计算限制,没有获得图像质量测量值。使用点源对所有扫描仪环配置进行了空间分辨率模拟,以检查视差误差的影响。使用标准 70cm 散射体和长度为 200cm 的扩展散射体对所有扫描仪进行了 NEMA 计数率估计,散射体中含有(1-800)MBq 的 F。使用 NEMA 方法对 70cm 标准和 200cm 长线源进行了灵敏度评估。

结果

在距扫描仪中心 1 和 10cm 处使用滤波反投影重建的径向/切向/轴向空间分辨率的平均值分别为 3.94/4.10/4.41mm 和 5.29/4.89/5.90mm对于四环扫描仪。使用 70cm 线源,在 FOV 中心处确定四环扫描仪的灵敏度为 14.86cps/kBq。与四环扫描仪相比,使用 200cm 长线源时,1 和 2m AFOV 扫描仪的灵敏度提高了 21 倍和 60 倍。在 2m AFOV 扫描仪中的空间分辨率模拟表明,与四环扫描仪相比,轴向分辨率最大下降约 23.8%。然而,当增加轴向接收角度到±70°时,横向分辨率被发现相对稳定。四环扫描仪的最大 NECR 为 212.92kcps,在 22.70kBq/ml 时的散射分数为 38.9%,使用 70cm 散射体。与四环扫描仪相比,使用 200cm 长散射体对峰值 NECR 的比较导致 20 环的 NECR 增益为 15,40 环的增益为 28。与发表的测量值相比,空间分辨率、灵敏度和散射分数的一致性在 7%以内。

结论

成功地将四环 DMI 扫描仪模拟与发表的 NEMA 测量值进行了比较。随后模拟了基于 DMI 架构的扩展 1 和 2m AFOV 扫描仪的灵敏度和 NECR 性能。与使用 Siemens mCT 架构扩展和发表的 NEMA 测量值的先前模拟研究以及使用 uEXPLORER 系统的出版物一致,灵敏度和计数率性能得到了提高。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验