Bieszk J A, Hawman E G
J Nucl Med. 1987 Aug;28(8):1308-14.
In order to understand the angular dependence of reconstructed images, a "hot rod" phantom was simulated with six rod sizes ranging from 12.7 mm to 4.8 mm, arranged in 60 degree sectors. This simulation was used to evaluate changes caused by increasing the number of views from 60 to 180 and to compare results to published angular sampling requirements. Also, a comparison was made of step-and-shoot and continuous data acquisition modes. Experimental phantom data were taken for three acquisition modes and for two levels of image statistics and compared with the simulation. This study shows the effects of angular aliasing and that image quality generally improves with increasing views even if total image counts are reduced. Despite smoothing effects, a continuous (180 view) scan is, in most cases, the preferred acquisition mode because it offers the highest angular sampling, as well as maximum counting efficiency.
为了了解重建图像的角度依赖性,使用“热棒”模型进行模拟,该模型有六种直径从12.7毫米到4.8毫米的棒,以60度扇形排列。此模拟用于评估将视图数量从60增加到180所引起的变化,并将结果与已发表的角度采样要求进行比较。此外,还对步进采集和连续数据采集模式进行了比较。针对三种采集模式以及两种图像统计水平采集了实验模型数据,并与模拟结果进行比较。本研究显示了角度混叠的影响,并且表明即使总图像计数减少,图像质量通常也会随着视图数量的增加而提高。尽管有平滑效果,但在大多数情况下,连续(180视图)扫描是首选的采集模式,因为它提供了最高的角度采样以及最大的计数效率。