Tekin F, Can-Akman T, Kitiş A
Pamukkale University, School of Physical Therapy and Rehabilitation, Çamlaraltı District, University Street, Kınıklı Campus, No: 11, 20160, Denizli, Turkey.
Hand Surg Rehabil. 2022 Apr;41(2):183-188. doi: 10.1016/j.hansur.2021.12.006. Epub 2021 Dec 30.
The aim of this study was to determine the validity and reliability of the KFORCE Sens® electrogoniometer in the evaluation of wrist proprioception. Wrist position sense was assessed on a Baseline® 360° universal goniometer and a KFORCE Sens® device. The validity and reliability of the KFORCE Sens® device for wrist position sense evaluation were investigated by comparing the two data sets. Fifty-three healthy volunteers (39 female, 14 male) with a mean age of 22.83 ± 1.28 years (range, 21-27 years) were included. Joint position sense test-retest reliability (intra-class correlation coefficient) on KFORCE Sens® was "very good" for all wrist movements. There was a very strong correlation between flexion-extension movements on the dominant side (r = 0.955), and a strong correlation between ulnar-radial deviation movements (r = 0.745). There was also a very strong (r = 0.863) correlation between flexion-extension movements on the non-dominant side and a strong correlation (r = 0.690) between ulnar-radial deviation movements (p < 0.05). Our results showed that the KFORCE Sens® device was a valid and reliable evaluation means of assessing wrist position sense.
本研究的目的是确定KFORCE Sens®电子测角仪在评估腕关节本体感觉方面的有效性和可靠性。在Baseline® 360°通用测角仪和KFORCE Sens®设备上评估腕关节位置觉。通过比较两组数据集,研究KFORCE Sens®设备用于评估腕关节位置觉的有效性和可靠性。纳入了53名健康志愿者(39名女性,14名男性),平均年龄为22.83±1.28岁(范围21 - 27岁)。KFORCE Sens®上的关节位置觉重测信度(组内相关系数)对于所有腕关节运动均为“非常好”。优势侧屈伸运动之间存在非常强的相关性(r = 0.955),尺桡偏斜运动之间存在强相关性(r = 0.745)。非优势侧屈伸运动之间也存在非常强的相关性(r = 0.863),尺桡偏斜运动之间存在强相关性(r = 0.690)(p < 0.05)。我们的结果表明,KFORCE Sens®设备是评估腕关节位置觉的一种有效且可靠的评估手段。