National Institute of Cardiovascular Diseases, Karachi, Pakistan.
National Institute of Cardiovascular Diseases (NICVD), Karachi.
J Pak Med Assoc. 2021 Oct;71(10):2461-2463. doi: 10.47391/JPMA.06-844.
We present an unusual case of positional syncope occurring years after pacemaker implantation. Syncope was caused by a pacemaker lead fracture, which resulted from subclavian crush syndrome. The syncope occurred incidentally during hospital admission and was timely diagnosed using an integrated approach of history taking, examination findings, device interrogation and radiographic parameters. The patient subsequently underwent lead and device alteration which led to resolution of her symptoms.
我们报告一例起搏器植入多年后发生的位置性晕厥。晕厥是由锁骨下压迫综合征引起的起搏器导线断裂导致的。晕厥是在住院期间偶然发生的,通过病史采集、检查结果、设备检测和影像学参数的综合方法及时诊断。随后,患者进行了导线和设备的更换,症状得到缓解。