Performance of an automated ultrasound device in identifying and tracing the heart in porcine cardiac arrest.

作者信息

Olszynski Paul, Marshall Rory A, Olver T Dylan, Oleniuk Trevor, Auser Cameron, Wilson Tracy, Atkinson Paul, Woods Rob

机构信息

Department of Emergency Medicine, University of Saskatchewan, 103 Hospital Drive, Saskatoon, SK, S7N 0W8, Canada.

Western College of Veterinary Medicine, University of Saskatchewan, Saskatoon, Canada.

出版信息

Ultrasound J. 2022 Jan 3;14(1):1. doi: 10.1186/s13089-021-00251-5.

Abstract

BACKGROUND

While intra-arrest echocardiography can be used to guide and monitor chest compression quality, it is not currently feasible on the scene of out-of-hospital cardiac arrests. Rapid and automated sonographic localization of the heart may provide first-responders guidance to an optimal area of compression without requiring them to interpret ultrasound images. In this proof-of-concept porcine study, we sought to describe the performance of an automated ultrasound device in correctly identifying and tracing the borders of the heart in three distinct states: pre-arrest, arrest, and late arrest.

METHODS

An automated ultrasound device (bladder scanner) was placed on the chests of 7 swine, along the left sternal border (4th-8th intercostal spaces). Scanner-generated images were recorded for each space during pre-arrest, arrest, and finally late arrest. 828 images of the LV and LV outflow tract were randomized and 150 (50/state) selected for analysis. Scanner tracings of the heart were then digitally obscured to facilitate tracing by expert reviewers who were blinded to the physiologic state. Reviewer tracings were compared to bladder scanner tracings; with concordance between these images determined via Sørensen-Dice index (SDI).

RESULTS

When compared to human reviewers, the bladder scanner was able to identify and trace the borders during cardiac arrest. The bladder scanner performed best at the time of arrest (SDI 0.900 ± 0.059). As resuscitation efforts continued and time from initial arrest increased, the scanner's performance decreased dramatically (SDI 0.597 ± 0.241 in late arrest).

CONCLUSION

An automated ultrasound device (bladder scanner) reliably traced porcine hearts during cardiac arrest. It is possible a device could be developed to indicate where compressions should be performed without requiring the operator to interpret ultrasound images. Further investigation into rapid, automated, sonographic localization of the heart to identify the area of compression in out-of-hospital cardiac arrest is warranted.

摘要

背景

虽然心脏骤停期间的超声心动图可用于指导和监测胸外按压质量,但目前在院外心脏骤停现场尚不可行。快速自动的心脏超声定位可为急救人员提供最佳按压区域的指导,而无需他们解读超声图像。在这项概念验证性猪研究中,我们试图描述一种自动超声设备在三种不同状态下(心脏骤停前、心脏骤停时和心脏骤停后期)正确识别和追踪心脏边界的性能。

方法

将一台自动超声设备(膀胱扫描仪)沿左胸骨旁线(第4 - 8肋间)放置在7头猪的胸部。在心脏骤停前、心脏骤停时以及最终心脏骤停后期,记录每个肋间空间由扫描仪生成的图像。随机选取828张左心室和左心室流出道的图像,其中150张(每种状态50张)用于分析。然后将心脏的扫描仪追踪图像进行数字模糊处理,以便由对生理状态不知情的专家审阅者进行追踪。将审阅者的追踪结果与膀胱扫描仪的追踪结果进行比较;通过 Sørensen-Dice 指数(SDI)确定这些图像之间的一致性。

结果

与人类审阅者相比,膀胱扫描仪能够在心脏骤停期间识别并追踪心脏边界。膀胱扫描仪在心脏骤停时表现最佳(SDI 为0.900 ± 0.059)。随着复苏努力的持续以及从最初心脏骤停开始的时间增加,扫描仪的性能显著下降(心脏骤停后期SDI 为0.597 ± 0.241)。

结论

一种自动超声设备(膀胱扫描仪)在心脏骤停期间能够可靠地追踪猪的心脏。有可能开发一种设备,无需操作人员解读超声图像就能指示应进行按压的位置。有必要进一步研究快速、自动的心脏超声定位,以确定院外心脏骤停时的按压区域。