Efficacy of MAVIG X-Ray Protective Drapes in Reducing CTO Operator Radiation.

机构信息

Department of Cardiovascular Medicine, University Hospitals Leuven, Leuven, Belgium.

Department of Cardiovascular Sciences, Katholieke Universiteit, Leuven, Belgium.

出版信息

J Interv Cardiol. 2021 Dec 14;2021:3146104. doi: 10.1155/2021/3146104. eCollection 2021.

Abstract

BACKGROUND

The MAVIG X-ray protective drape (MXPD) has been shown to reduce operator radiation dose during percutaneous coronary interventions (PCI). Whether MXPDs are also effective in reducing operator radiation during chronic total occlusion (CTO) PCI, often with dual access, is unknown.

METHODS

We performed a prospective, randomized-controlled study comparing operator radiation dose during CTO PCI ( = 60) with or without pelvic MXPDs. The primary outcomes were the difference in first operator radiation dose (Sv) and relative dose of the first operator (radiation dose normalized for dose area product) at the level of the chest in the two groups. The effectiveness of MXPD in CTO PCI was compared with non-CTO PCI using a patient-level pooled analysis with a previously published non-CTO PCI randomized study.

RESULTS

The use of the MXPD was associated with a 37% reduction in operator dose (weighted median dose 26.0 (IQR 10.00-29.47) Sv in the drape group versus 41.8 (IQR 30.82-60.59) Sv in the no drape group; < 0.001) and a 60% reduction in relative operator dose (median dose 3.5 (IQR 2.5-5.4) E/DAPx10 in the drape group versus 8.6 (IQR 4.2-12.5) E/DAPx10 in the no drape group; =0.001). MXPD was equally effective in reducing operator dose in CTO PCI compared with non-CTO PCI ( value for interaction 0.963).

CONCLUSIONS

The pelvic MAVIG X-ray protective drape significantly reduced CTO operator radiation dose. This trial is clinically registered with https://www.clinicaltrials.gov (unique identifier: NCT04285944).

摘要

背景

MAVIG X 射线防护帘(MXPD)已被证明可降低经皮冠状动脉介入治疗(PCI)期间术者的辐射剂量。在慢性完全闭塞(CTO)PCI 中,尤其是双入路时,MXPD 是否也能有效降低术者的辐射剂量尚不清楚。

方法

我们进行了一项前瞻性、随机对照研究,比较了 CTO PCI( = 60)中使用或不使用骨盆 MXPD 时术者的辐射剂量。主要结局是两组术者胸部水平的第一术者辐射剂量(Sv)和第一术者相对剂量(剂量面积乘积归一化的辐射剂量)的差异。使用先前发表的非 CTO PCI 随机研究的患者水平汇总分析,比较了 MXPD 在 CTO PCI 中的有效性。

结果

使用 MXPD 可使术者剂量降低 37%(覆盖组的加权中位数剂量为 26.0(IQR 10.00-29.47)Sv,无覆盖组为 41.8(IQR 30.82-60.59)Sv; < 0.001),术者相对剂量降低 60%(覆盖组中位数剂量为 3.5(IQR 2.5-5.4)E/DAPx10,无覆盖组为 8.6(IQR 4.2-12.5)E/DAPx10;=0.001)。与非 CTO PCI 相比,MXPD 在降低 CTO PCI 术者剂量方面同样有效(交互值为 0.963)。

结论

骨盆 MAVIG X 射线防护帘可显著降低 CTO 术者的辐射剂量。该试验已在 https://www.clinicaltrials.gov 注册(唯一标识符:NCT04285944)。