Sato S, Mimura H, Noda T, Orita K, Hiraki Y, Awai S, Nagaya I, Aono K
1st Department of Surgery, Okayama University Medical School.
Radioisotopes. 1987 Jun;36(6):265-9. doi: 10.3769/radioisotopes.36.6_265.
The quantitative evaluation on measurement of liver volume and the hepatic accumulation rate of the radioactive agent were studied with respect to body constitution, background activity and splenic consumption rate, using single photon emission computed tomography (SPECT). The optical cut-off level for liver volume determination was proved to be reasonable as 35%. The background less than 5% had no influence to their clinical estimation with the error of less than 1.6%. The estimation of liver volume was mostly influenced by the splenic activity. When the splenic activity was increased two times of the hepatic activity, the liver volume was determined as 4.2% larger. Calculated counts by SPECT was well correlated with the hepatic activity. This study proved that SPECT was useful for the quantitative evaluation of the liver volume and the hepatic accumulation rate.
使用单光子发射计算机断层扫描(SPECT),针对身体体质、本底活性和脾脏摄取率,对肝脏体积测量及放射性药物的肝脏摄取率进行了定量评估。确定肝脏体积的光学截止水平为35%被证明是合理的。本底低于5%对临床评估无影响,误差小于1.6%。肝脏体积的评估主要受脾脏活性影响。当脾脏活性增加至肝脏活性的两倍时,肝脏体积测定值大4.2%。SPECT计算的计数与肝脏活性高度相关。本研究证明SPECT对于肝脏体积和肝脏摄取率的定量评估是有用的。