T1 加权 ihMT 成像 - 第二部分。研究长 T 成分和短 T 成分与髓鞘含量的相关性。与 R 和大分子质子分数的比较。

T -weighted ihMT imaging - Part II. Investigating the long- and short-T components correlation with myelin content. Comparison with R and the macromolecular proton fraction.

机构信息

Aix Marseille Univ, CNRS, CRMBM, Marseille, France.

APHM, Hôpital Universitaire Timone, CEMEREM, Marseille, France.

出版信息

Magn Reson Med. 2022 May;87(5):2329-2346. doi: 10.1002/mrm.29140. Epub 2022 Jan 9.

Abstract

PURPOSE

To investigate the long- and short-T components correlation with myelin content using inhomogeneous magnetization transfer (ihMT) high-pass and band-pass T -filters and to compare ihMT, R , and the macromolecular proton fraction (MPF) for myelin specific imaging.

METHODS

The 3D ihMT rapid gradient echo (ihMTRAGE) sequences with increasing switching times (Δt) were used to derive ihMT high-pass T -filters with increasing T cutoff values and an ihMT band-pass T -filter for components in the 100 µs to 1 ms range. 3D spoiled gradient echo quantitative MT (SPGR-qMT) protocols were used to derive R and MPF maps. The specificity of R , MPF, and ihMT T -filters was evaluated by comparison with two histological reference techniques for myelin imaging.

RESULTS

The higher contribution of long-T s as compared to the short components as Δt got longer led to an increase in the specificity to myelination. In contrast, focusing on the signal originating from a narrow range of short-T s (< 1 ms) as isolated by the band-pass T -filter led to lower specificity. In addition, the significantly lower r correlation coefficient of the band-pass T -filter suggests that the origin of short-T components is mostly associated with non-myelin protons. Also, the important contribution of short-T s to the estimated MPF, explains its low specificity to myelination as compared to the ihMT high-pass T -filters.

CONCLUSION

Long-T components imaging by means of ihMT high-pass T -filters is proposed as an MRI biomarker for myelin content. Future studies should enable the investigation of the sensitivity of ihMT T -filters for demyelinating processes.

摘要

目的

利用不均匀磁化传递(ihMT)高通和带通 T 滤波器研究长 T 和短 T 成分与髓鞘含量的相关性,并比较 ihMT、R 和大分子质子分数(MPF)在髓鞘特异性成像中的应用。

方法

采用 3D ihMT 快速梯度回波(ihMTRAGE)序列,增加切换时间(Δt),得到具有不同 T 截止值的 ihMT 高通 T 滤波器和 100μs 到 1ms 范围内的 ihMT 带通 T 滤波器。采用 3D 扰相梯度回波定量 MT(SPGR-qMT)方案得出 R 和 MPF 图。通过与两种用于髓鞘成像的组织学参考技术进行比较,评估 R、MPF 和 ihMT T 滤波器的特异性。

结果

随着Δt 的增加,长 T 成分的贡献相对于短 T 成分增加,导致髓鞘成像的特异性提高。相反,通过带通 T 滤波器聚焦于短 T 成分(<1ms)的信号,会导致特异性降低。此外,带通 T 滤波器的 r 相关系数显著降低,表明短 T 成分的起源主要与非髓鞘质子有关。此外,短 T 成分对估计 MPF 的重要贡献解释了其与 ihMT 高通 T 滤波器相比对髓鞘化的特异性较低。

结论

建议采用 ihMT 高通 T 滤波器对长 T 成分进行成像,作为髓鞘含量的 MRI 生物标志物。未来的研究应该能够研究 ihMT T 滤波器对脱髓鞘过程的敏感性。