Sanju S, Jain Paresh, Priya Veeraraghavan Vishnu, Varma Praveen K, Mony Ullas
Centre for Nanosciences and Molecular Medicine, Amrita VishwaVidyapeetham, Kochi, 682041, Kerala, India.
BD Biosciences, Singapore, Singapore.
Appl Biochem Biotechnol. 2023 Sep;195(9):5747-5752. doi: 10.1007/s12010-022-03819-6. Epub 2022 Jan 11.
Quantitation of mHLA-DR and nCD64 is useful in understanding the dysregulated host response. The down regulation of HLA-DR expression on the circulating monocytes (mHLA-DR) is associated with anti-inflammatory response, and an increased expression of CD64 on neutrophil surface (nCD64) is associated with pro-inflammatory response. Quantitation of these antigen expression using beads (QuantiBRITE™ PE) is a precision technique. These beads are reported to be stable for 24 h after reconstitution. We report the results of our investigation examining the stability of QuantiBRITE PE beads over a period of 4-week post-reconstitution. The data suggest that reconstituted QuantiBRITE PE beads, if stored in dark at 2-8 °C, can be effectively used for up to 2 weeks for determining nCD64 and mHLA-DR antibody bound per cell (ABC) values.