Elkhateeb Sara Mohamed, Aloyouny Ashwag Yagoub, Omer Mohamed Mohamed Saeed, Mansour Soad Mahmoud
Department of Basic Dental Science, College of Dentistry, Princess Nourah bint Abdulrahman University, Riyadh 11671, Saudi Arabia.
College of Dentistry, Princess Nourah bint Abdulrahman University, Riyadh 11671, Saudi Arabia.
World J Clin Cases. 2022 Jan 14;10(2):437-447. doi: 10.12998/wjcc.v10.i2.437.
Digital radiography has recently been used in dentistry as a substitute for conventional film radiography worldwide. Digital imaging has many advantages and provides new possibilities for recording and interpreting radiographic data. This system uses different types of digital receptors.
To detect the frequency, type, and reasons behind the appearance of intraoral image artifacts acquired by photostimulable phosphor plates (PSP).
This retrospective descriptive study was conducted in the oral and maxillofacial radiology unit of the dental clinics of the College of Dentistry, Princess Nourah University (PNU). All intraoral digital radiographs were acquired using (Gendex Expert DC., United States) an intraoral X-ray machine with 7 -mA, 65-kVP using a PSP system (Soredex DIGORA Optime imaging plate) and laser scanners (Soredex DIGORA Optime), which can house all sizes of reusable intraoral PSP sensor plates with image acquisition software (MIPACS Dental Enterprise viewer 3.2.2). A total of 50000 intraoral radiographs were retrieved from the clinical database from April 2018 to April 2020 to evaluate the reason, type, and solutions to these image artifacts.
Overall, 50000 intraoral digital radiographs were acquired in a two-year-period; that is, from April 2018 to April 2020. Of these, 3550 (7.1%) retakes were performed due to the presence of image artifacts. Periapical radiographs were the most common image type of intraoral retakes (80.8%). Imaging artifacts were divided into three categories: operator, plate and scanning errors. Out of 3550 retakes, 5%, 1.37%, and 0.73% were related to operator, plate, and scanning errors, respectively. The cone cut was the most common operator error (988 images), Bite marks were the most common plate error (276 images), and delayed scanning artifacts were the most common scanning errors (145 images). The calculated kappa value for interobserver reliability was 0.99, indicating almost perfect interobserver agreement.
Our study discussed intraoral image artifacts that were characteristic of PSP, where the most common artifacts were bitemarks, image size reduction, scratches, and delayed scanning.
数字放射成像最近在全球范围内已被用于牙科,以替代传统的胶片放射成像。数字成像具有许多优点,并为记录和解读放射影像数据提供了新的可能性。该系统使用不同类型的数字接收器。
检测光激励荧光板(PSP)获取的口腔内图像伪影出现的频率、类型及原因。
本回顾性描述性研究在努拉公主大学牙科学院牙科诊所的口腔颌面放射科进行。所有口腔内数字放射照片均使用(美国Gendex Expert DC.)一台7毫安、65千伏峰值的口腔内X射线机,采用PSP系统(Soredex DIGORA Optime成像板)和激光扫描仪(Soredex DIGORA Optime)获取,该激光扫描仪可容纳各种尺寸的可重复使用口腔内PSP传感器板以及图像采集软件(MIPACS Dental Enterprise viewer 3.2.2)。从2018年4月至2020年4月的临床数据库中检索了总共50000张口腔内放射照片,以评估这些图像伪影的原因、类型及解决方法。
总体而言,在两年期间,即从2018年4月至2020年4月,共获取了50000张口腔内数字放射照片。其中,由于存在图像伪影,进行了3550次(7.1%)重新拍摄。根尖片是口腔内重新拍摄中最常见的图像类型(80.8%)。成像伪影分为三类:操作者、成像板和扫描错误。在3550次重新拍摄中,分别有5%、1.37%和0.73%与操作者、成像板和扫描错误有关。锥形束投照失误是最常见的操作者错误(988张图像),咬痕是最常见的成像板错误(276张图像),延迟扫描伪影是最常见的扫描错误(145张图像)。观察者间可靠性的计算kappa值为0.99,表明观察者间几乎完全一致。
我们的研究讨论了PSP特有的口腔内图像伪影,其中最常见的伪影是咬痕、图像尺寸缩小、划痕和延迟扫描。