吞咽困难患者的吞咽评估:一种袖珍超声系统的有效性和可靠性。

Swallowing assessment in patients with dysphagia: Validity and reliability of a pocket-sized ultrasound system.

作者信息

Winiker Katharina, Hammond Rebecca, Thomas Paige, Dimmock Alice, Huckabee Maggie-Lee

机构信息

Department of Psychology, Speech and Hearing, University of Canterbury, Christchurch, New Zealand.

Rose Centre for Stroke Recovery and Research, University of Canterbury, Christchurch, New Zealand.

出版信息

Int J Lang Commun Disord. 2022 May;57(3):539-551. doi: 10.1111/1460-6984.12703. Epub 2022 Feb 3.

Abstract

BACKGROUND

The use of ultrasound as an adjunct to clinical swallowing evaluation provides quantitative physiological and morphological data. As a low-risk procedure, ultrasound imaging can be performed outside of a medical setting. This is particularly important for patients living in rural areas with restricted access to a hospital. Technical advances have produced pocket-sized ultrasound technology that is more affordable, and therefore within the fiscal reach of most allied health services.

AIMS

To explore the validity and reliability of pocket-sized ultrasound technology in dysphagia assessment.

METHODS & PROCEDURES: Data were acquired from 43 patients with dysphagia using the Clarius ultrasound device. Ultrasound and videofluoroscopic measures of hyoid and laryngeal displacement during liquid and puree swallowing were collected concurrently to quantify correlation and agreement between identical measures derived from the two instruments. Reliability of ultrasound was assessed for measures of hyoid and laryngeal displacement, tongue thickness, and size of the submental muscles in eight patients. Reliability was evaluated for the entire process of data acquisition including scanning and online measurement using an iPad in a clinical setting and for offline measurement on a computer screen to explore environmental influences on reliability.

OUTCOMES & RESULTS: Results revealed poor correlation between the measures of interest across instruments. Reliability of the entire process of data acquisition in a clinical setting was insufficient while reliability was more promising for offline measurements.

CONCLUSIONS & IMPLICATIONS: The clinical use of pocket-sized ultrasound devices, such as the Clarius system, for swallowing evaluation is not indicated at this time. Enhanced validity and reliability of the entire process of data acquisition are needed prior to clinical translation of such technology.

WHAT THIS PAPER ADDS

What is already known on the subject The use of ultrasound allows for radiation-free, non-invasive swallowing assessment. Some data suggest that ultrasound is valid and reliable in the evaluation of swallowing using standard-sized equipment. Insufficient validity and reliability have been reported for pocket-sized ultrasound technology in the assessment of healthy swallowing. What this paper adds to existing knowledge This research is the first to provide validity and reliability data of the pocket-sized Clarius technology in the evaluation of swallowing in patients with dysphagia. Insufficient validity and reliability of online data acquisition in a clinical environment were found. Reliability for offline measurement was more promising. What are the potential or actual clinical implications of this work? The clinical use of pocket-sized ultrasound devices, such as the Clarius system, for swallowing assessment is not indicated at this time.

摘要

背景

超声作为临床吞咽评估的辅助手段,可提供定量的生理和形态学数据。作为一种低风险的检查方法,超声成像可在医疗场所之外进行。这对于居住在农村地区、就医受限的患者尤为重要。技术进步催生了袖珍型超声技术,其价格更为亲民,因此大多数联合健康服务机构在财务上均可负担。

目的

探讨袖珍型超声技术在吞咽障碍评估中的有效性和可靠性。

方法与步骤

使用Clarius超声设备从43例吞咽障碍患者处获取数据。在液体和泥状食物吞咽过程中,同时收集舌骨和喉部位移的超声及视频荧光透视测量数据,以量化两种仪器相同测量结果之间的相关性和一致性。对8例患者的舌骨和喉部位移、舌厚度及颏下肌肉大小的超声测量可靠性进行评估。对包括在临床环境中使用iPad进行扫描和在线测量以及在电脑屏幕上进行离线测量的数据采集全过程的可靠性进行评估,以探究环境因素对可靠性的影响。

结果

结果显示不同仪器间相关测量指标的相关性较差。临床环境下数据采集全过程的可靠性不足,而离线测量的可靠性更具前景。

结论与启示

目前不建议临床使用袖珍型超声设备(如Clarius系统)进行吞咽评估。在此类技术临床应用之前,需要提高数据采集全过程的有效性和可靠性。

本文的新增内容

关于该主题的已知信息 超声检查可实现无辐射、非侵入性的吞咽评估。一些数据表明,使用标准尺寸设备时,超声在吞咽评估中是有效且可靠的。在健康吞咽评估中,袖珍型超声技术的有效性和可靠性不足。本文对现有知识的补充 本研究首次提供了袖珍型Clarius技术在吞咽障碍患者吞咽评估中的有效性和可靠性数据。发现临床环境下在线数据采集的有效性和可靠性不足。离线测量的可靠性更具前景。这项工作的潜在或实际临床意义是什么?目前不建议临床使用袖珍型超声设备(如Clarius系统)进行吞咽评估。