关于从聚氨酯泡沫中标准化拧出螺钉的建议 - 测试泡沫密度变化和插入方法的影响。

Recommendations for standardised screw pull-out from polyurethane foam - The influence of density variations of the test foam and the insertion method.

机构信息

ZESBO - Center for Research on Musculoskeletal Systems, Department of Orthopaedic Surgery, Traumatology and Plastic Surgery, University of Leipzig, Leipzig, Germany.

ZESBO - Center for Research on Musculoskeletal Systems, Department of Orthopaedic Surgery, Traumatology and Plastic Surgery, University of Leipzig, Leipzig, Germany.

出版信息

Med Eng Phys. 2022 Feb;100:103750. doi: 10.1016/j.medengphy.2022.103750. Epub 2022 Jan 2.

Abstract

Insertion and pull-out tests of synthetic test material are well established for the initial laboratory evaluation of screws. However, not all test parameters are sufficiently described. The influence of small density deviations of the test material, of tapping or of manual or machine insertion has not been fully examined. The aim of the present study was to examine the influence of these specimen preparation parameters on the measurement results in order to increase the reproducibility and reliability of screw pull-out tests. For this purpose, a commercial polyurethane foam and a clinically used type of screw are evaluated with insertion and pull-out tests. Within a foam apparent density grade, small deviations in apparent density led to significant and relevant differences in the measured values of insertion torque and pull-out strength. Furthermore, an influence on the measurement results was found during tapping and during manual or machine insertion of screws. For these reasons, specimens with the same apparent density should be used as far as possible and evenly distributed among the test groups. In addition, the reproducibility of the results can be increased by machine insertion of the screws.

摘要

对于螺钉的初步实验室评估,合成测试材料的插入和拔出测试已经得到很好的确立。然而,并非所有测试参数都得到了充分描述。测试材料的小密度偏差、攻丝或手动或机器插入的影响尚未得到充分检查。本研究的目的是检查这些样品制备参数对测量结果的影响,以提高螺钉拔出测试的可重复性和可靠性。为此,使用插入和拔出测试评估了一种商业聚氨酯泡沫和一种临床使用的螺钉类型。在泡沫表观密度等级内,表观密度的微小偏差导致插入扭矩和拔出强度测量值的显著和相关差异。此外,在攻丝以及手动或机器插入螺钉时,发现对测量结果有影响。因此,应尽可能使用具有相同表观密度的试件,并在测试组中均匀分布。此外,通过机器插入螺钉可以提高结果的可重复性。

相似文献

2
A novel parameter for the prediction of pedicle screw fixation in cancellous bone - A biomechanical study on synthetic foam.
Med Eng Phys. 2020 May;79:44-51. doi: 10.1016/j.medengphy.2020.03.001. Epub 2020 Mar 17.
3
Characteristics of immediate and fatigue strength of a dual-threaded pedicle screw in cadaveric spines.
Spine J. 2013 Aug;13(8):947-56. doi: 10.1016/j.spinee.2013.03.010. Epub 2013 Apr 17.
4
Experimental investigation of pull-out performance of pedicle screws at different polyurethane (PU) foam densities.
Proc Inst Mech Eng H. 2021 Jun;235(6):709-716. doi: 10.1177/09544119211002587. Epub 2021 Mar 17.
5
Pedicle screw pull-out testing in polyurethane foam blocks: Effect of block orientation and density.
Proc Inst Mech Eng H. 2024 Apr;238(4):455-460. doi: 10.1177/09544119241236873. Epub 2024 Mar 13.
8
Effect of various factors on pull out strength of pedicle screw in normal and osteoporotic cancellous bone models.
Med Eng Phys. 2017 Feb;40:28-38. doi: 10.1016/j.medengphy.2016.11.012. Epub 2016 Dec 9.
9
Screw pull-out force is dependent on screw orientation in an anterior cervical plate construct.
J Spinal Disord Tech. 2007 Jul;20(5):369-73. doi: 10.1097/BSD.0b013e31802c2a4a.
10
Feasibility of using tapping torque during lumbar pedicle screw insertion to predict screw fixation strength.
J Orthop Sci. 2020 May;25(3):389-393. doi: 10.1016/j.jos.2019.05.009. Epub 2019 Jun 4.