Xing Y, Hong M
Nanjing Prevention and Treatment Center for Occupational Diseases, Nanjing 210042, China.
Zhonghua Lao Dong Wei Sheng Zhi Ye Bing Za Zhi. 2021 Dec 20;39(12):947-949. doi: 10.3760/cma.j.cn121094-20200519-00274.
This study was aimed to evaluate the accuracy and conformity of the dry chemical urine analyzer UA-5800 and the urine formed element analyzer EH-2080Cused in the detection of urine red blood cells, and establish screening and retesting standards for urine red blood cell testing. Automatic urine analysis line EU-8000 (composed of dry chemical urine analyzer UA-5800 and urine formed element analyzer EH-2080C) was used to conduct hematuria and urine erythrocyte test in 2336 fresh urine samples from occupational medical examinations in our hospital, the sensitivity (true positive rate) , specificity (true negative rate) and the positive coincidence rate (positive rate) and negative coincidence rate (negative rate) of the corresponding test results were analyzed tow instruments. The sensitivity of urinary occult blood detected by UA-5800 and urinary erythrocytes examined by EH-2080C were 98.33% and 95.00%, there was no statistical significant between the two results ((2)=0.50、>0.05) ; specificity was 82.47% and 97.45% respectively, which existed statistical significant ((2)=337.05、<0.01) . Comparing the two methods, the statistical significant remained ((2)=339.05、<0.01) . Among all results of urinary occult blood detected, the EH-2080C had the highest urinary erythrocyte negative coincidence rate (99.89%) ; while in the comparison of urinary occult blood positive grades at 3+, the highest positive coincidence rate of urinary erythrocyte in EH-2080C was 85.76%; when the urine occult blood levels were weakly positive, 1+, 2+, the positive coincidence rate of EH-2080C examination was low (8.97%, 14.40% and 36.48%) , which increased with elevated positive grade. The positive specimens screened by automatic urine analysis line were examined by gold standard microscope combining with with the criteria of urine red blood cell screening and retesting in our laboratory, which can ensure high accuracy and high efficiency of urine erythrocyte analysis.
本研究旨在评估干化学尿液分析仪UA - 5800和尿液有形成分分析仪EH - 2080C用于检测尿红细胞的准确性和一致性,并建立尿红细胞检测的筛查和复检标准。采用自动尿液分析流水线EU - 8000(由干化学尿液分析仪UA - 5800和尿液有形成分分析仪EH - 2080C组成)对我院职业健康检查的2336份新鲜尿液样本进行血尿和尿红细胞检测,分析两台仪器相应检测结果的灵敏度(真阳性率)、特异度(真阴性率)、阳性符合率(阳性率)和阴性符合率(阴性率)。UA - 5800检测尿隐血和EH - 2080C检测尿红细胞的灵敏度分别为98.33%和95.00%,两者结果差异无统计学意义((2)=0.50、>0.05);特异度分别为82.47%和97.45%,差异有统计学意义((2)=337.05、<0.01)。两种方法比较,差异仍有统计学意义((2)=339.05、<0.01)。在所有尿隐血检测结果中,EH - 2080C的尿红细胞阴性符合率最高(99.89%);在尿隐血阳性等级为3 +的比较中,EH - 2080C尿红细胞的最高阳性符合率为85.76%;当尿隐血水平为弱阳性、1 +、2 +时,EH - 2080C检测的阳性符合率较低(8.97%、14.40%和36.48%),随阳性等级升高而升高。自动尿液分析流水线筛查出的阳性标本采用金标准显微镜结合本实验室尿红细胞筛查和复检标准进行检查,可确保尿红细胞分析的高准确性和高效率。