PATIENT AND STAFF DOSES FOR VARIOUS INTERVENTIONAL RADIOLOGY AND CARDIOLOGY EXAMINATIONS IN TURKEY.

作者信息

Erdem Onur, Ay Musa, Yalcin Asena, Bilgic Sadık, Sanlıdilek Umman, Amasyalı Basri, Sancak Tanzer, Olgar Turan

机构信息

Ankara University, Institute of Nuclear Sciences, 06100 Ankara, Turkey.

Ankara University, Faculty of Medicine, Department of Radiology, 06100 Ankara, Turkey.

出版信息

Radiat Prot Dosimetry. 2022 Mar 16;198(3):158-166. doi: 10.1093/rpd/ncac006.

Abstract

This study aims to determine the radiation doses of patients and staff during different interventional radiology and cardiology examinations. Dose measurements for interventional radiology examinations were performed in Ibn-i Sina Hospital of Ankara University using Siemens Artis-Zee medical imaging system. Patient dose measurement was carried out for interventional cardiology examinations in Cardiology Department of TOBB-ETU University, Medical Faculty Hospital using Philips Allura Centron interventional X-ray system. Patient doses were obtained in terms of kerma area product (KAP) and cumulative air kerma (CAK) from KAP meter attached to the angiography system. Performance tests of the angiography system were performed before patient dose measurements. Staff dose measurements were carried out with thermoluminescence dosimeters (TLD-100) placed in certain areas on the staff. Patient dose measurements were performed for 15 different interventional radiology examinations on a total of 431 patients and for four different cardiology examinations on a total of 299 patients. Monte Carlo based PCXMC 2.0 program was used to calculate patient effective doses. Lower extremity arteriography was the most common examination with a mean KAP value of 30 Gy cm2 and mean effective dose value of 1.2 mSv for total number of 194 patients. Mean KAP values calculated for coronary angiography, percutaneous coronary intervention, electrophysiological procedures and radiofrequency cardiac ablation examinations were 62.8, 162.8, 16.7 and 70.6 Gy cm2, respectively. Radiologist, nurse and technician effective dose normalised to the unit KAP of patient dose were 0.15, 0.11 and 0.14 μSv Gy-1 cm-2. Similarly, cardiologist, nurse and technician effective dose normalised to the unit KAP of patient dose were 0.22, 0.15 and 0.09 μSv Gy-1 cm-2. Measured KAP and CAK values vary depending on the type and complexity of the examination. The measured staff doses during cardiac examinations were higher when compared with that measured for interventional radiology as expected.

摘要

本研究旨在确定不同介入放射学和心脏病学检查期间患者及工作人员所接受的辐射剂量。在安卡拉大学伊本·西那医院使用西门子Artis-Zee医学成像系统对介入放射学检查进行剂量测量。在TOBB-ETU大学医学院附属医院心内科使用飞利浦Allura Centron介入X射线系统对介入心脏病学检查进行患者剂量测量。通过连接到血管造影系统的剂量面积乘积(KAP)仪获取患者剂量的KAP和累积空气比释动能(CAK)。在患者剂量测量之前对血管造影系统进行性能测试。工作人员剂量测量通过放置在工作人员特定部位的热释光剂量计(TLD-100)进行。对总共431例患者进行了15种不同的介入放射学检查的患者剂量测量,对总共299例患者进行了4种不同的心脏病学检查的患者剂量测量。使用基于蒙特卡罗方法的PCXMC 2.0程序计算患者有效剂量。下肢动脉造影是最常见的检查,194例患者的平均KAP值为30 Gy cm²,平均有效剂量值为1.2 mSv。冠状动脉造影、经皮冠状动脉介入治疗、电生理手术和射频心脏消融检查的平均KAP值分别为62.8、162.8、16.7和70.6 Gy cm²。以患者剂量的单位KAP归一化后的放射科医生、护士和技术人员有效剂量分别为0.15、0.11和0.14 μSv Gy⁻¹ cm⁻²。同样,以患者剂量的单位KAP归一化后的心脏病专家、护士和技术人员有效剂量分别为0.22、0.15和0.09 μSv Gy⁻¹ cm⁻²。测量的KAP和CAK值因检查类型和复杂性而异。正如预期的那样,心脏检查期间测量的工作人员剂量高于介入放射学检查时测量的剂量。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验