Otolaryngol Pol. 2021 Sep 7;75(6):1-10. doi: 10.5604/01.3001.0015.2575.
<br><b>Introduction:</b> Advances in computer image analysis have enabled the use of new functional imaging methods in the diagnosis of laryngeal diseases. Particularly interesting techniques of dynamic laryngeal imaging involve High Speed Videoendoscopy (HSV). This still-developed technique allows to overcome the limitations of laryngovideostroboscopy (LVS) and a more detailed analysis of the glottal function based on the image of the actual vibrations of the vocal folds. It also enables the determination of objective coefficients parameterizing phonatory vibrations of the vocal folds.</br> <br><b>Aim:</b> The aim of this pilot study was to evaluate the use of a high-speed videoendoscopy set with laser illumination for the diagnosis of glottic pathology in ENT practice.</br> <br><b>Material and methods:</b> The study included 40 patients who underwent LVS followed by HSV. The modern HSV examination kit - Advanced Larynx Imager System (ALIS), used for the first time in a clinical setting in Poland, is characterized by significantly improved, compared to the previously used high-speed cameras, operational parameters - a light head, the possibility of continuous lighting operation without excessive heating of the head tip, registration of the image in full color scale. Thanks to such modernization, the safety and course of the examination do not differ from laryngoscopy conducted with commonly used recorders. The device owes some of these improvements to a laser illuminator which was used for the first time as the main light source in a high-speed camera. In the study, two cases were selected to present the results of HSV and the analysis of the generated kymograms - a woman with no glottic pathology and a man with a polyp of the right vocal fold. In the first case, the HSV examination compared with the LVS revealed a discrete glottis functional disorder in the form of a tendency to hyperphonation. The patient with an organic lesion had a clearly visible irregularity of vocal fold vibrations, which also allowed to trace mucosal wave disturbances related to its reflection from the pathological structure of the glottis and the formation of a return wave, both on the fold affected by the lesion and, to a lesser extent, contralaterally. The glottic dysfunctions observed in the studied patients were confirmed in the generated kymograms and the graphs of the glottal width waveform (GWW), as well as in the parameters calculated on their basis, assessing the frequency and amplitude of phonatory vibrations.</br> <br><b>Conclusions:</b> The use of high-speed videoendoscopy allows for a much more accurate assessment of the phonatory function of the glottis than in laryngovideostroboscopy. The presented HSV system allows for obtaining high quality kinematic images of the larynx, color fidelity, and contrast. The use of this technology in laryngological practice enables precise structural and functional assessment of the glottis and detection of discrete phonation disorders that elude the techniques used so far.</br>.
引言:计算机图像分析的进步使得在诊断喉病时可以使用新的功能成像方法。特别有趣的动态喉成像技术包括高速视频内窥镜(HSV)。这项仍在发展中的技术克服了喉视频频闪镜(LVS)的局限性,并基于声带实际振动的图像对声带功能进行更详细的分析。它还能够确定参数化声带发声振动的客观系数。
目的:本初步研究旨在评估在耳鼻喉科实践中使用带激光照明的高速视频内窥镜套件诊断声门疾病。
材料与方法:该研究纳入了 40 名接受 LVS 检查后再接受 HSV 检查的患者。该研究中首次在波兰临床环境中使用的现代 HSV 检查套件——高级喉成像系统(ALIS),与之前使用的高速摄像机相比,具有显著改进的操作参数——光头,可在不使探头尖端过热的情况下连续照明操作,以及全彩色图像记录。由于这种现代化,检查的安全性和过程与使用常用记录器进行的喉镜检查没有区别。该设备的一些改进归功于激光照明器,这是该设备首次作为高速摄像机的主要光源使用。在研究中,选择了两个案例来展示 HSV 的结果和生成的声谱图分析——一个无声门病变的女性和一个右侧声带息肉的男性。在第一个案例中,与 LVS 相比,HSV 检查显示出一种离散的声门功能障碍,表现为过度发声的趋势。患有器质性病变的患者的声带振动明显不规则,这也允许追踪与病变声门的粘膜波干扰有关的返回波,既在受病变影响的声带,也在较小程度上在对侧。在研究中观察到的声门功能障碍在生成的声谱图和声门宽度波形(GWW)的图表中得到了证实,以及基于这些图表计算的评估发声振动频率和幅度的参数中得到了证实。
结论:高速视频内窥镜的使用可以比喉视频频闪镜更准确地评估声门的发声功能。所展示的 HSV 系统允许获得高质量的喉部运动图像、色彩保真度和对比度。在喉镜实践中使用这项技术能够精确评估声门的结构和功能,并发现迄今为止使用的技术无法发现的离散发声障碍。