Trindade Hugo, Barreira Marisa, Xavier Cláudia
IQI-Instituto para a Qualidade em Imagem e Proteção Radiológica S.A., Lisboa, Portugal.
J Radiol Prot. 2022 Mar 21;42(2). doi: 10.1088/1361-6498/ac591b.
In Portugal, computed tomography (CT) exposures for adult routine head, adult routine abdomen and paediatric abdomen (5 years old) were evaluated between 2019 and 2020 in 92 CT units. The third quartile value for CTDIwas 63.2 mGy for adult head exams, 15.9 mGy for adult abdomen and 5.2 mGy for paediatric abdomen. However, the CTDIfor head examination in 7.6% of units exceeded the suspension level proposed in EU report 162 while the number of units exceeding suspension levels for adult and paediatric abdomen were 3.2% and 0.0%, respectively. The used acquisition parameters and the referred CT unit's specifications were studied to understand how radiation dose relates with tube voltage, tube current, rotation time, detector width, slice width, pitch value and the use of automatic exposure control (AEC). The axial acquisition mode was predominant (67%) in the head exams. However, the helical scanning was commonly used in abdomen exams. Dose reduction from adult to paediatric exams was achieved using a lower tube current (a 60% average reduction). AEC is preferably used in abdomen protocols (84% in adult exams and 96% in paediatric exams). Nonetheless, in adult routine head protocols only 27% used this system. Data was gathered from clinics and hospitals covered by this study which allowed an estimate of an annual frequency of 206 CT examinations per 1000 population in Portugal. The estimated dose index (CTDI) and frequency of CT examinations for the Portuguese population agreed with data from other countries.
2019年至2020年期间,在葡萄牙的92个CT设备中对成人常规头部、成人常规腹部和儿科腹部(5岁)的计算机断层扫描(CT)照射剂量进行了评估。成人头部检查的CTDI第三四分位数为63.2 mGy,成人腹部为15.9 mGy,儿科腹部为5.2 mGy。然而,7.6%的设备头部检查的CTDI超过了欧盟第162号报告中提出的暂停水平,而成人腹部和儿科腹部超过暂停水平的设备数量分别为3.2%和0.0%。研究了所使用的采集参数以及相关CT设备的规格,以了解辐射剂量与管电压、管电流、旋转时间、探测器宽度、层厚、螺距值以及自动曝光控制(AEC)的使用之间的关系。头部检查中轴向采集模式占主导(67%)。然而,腹部检查通常使用螺旋扫描。通过降低管电流(平均降低60%)实现了从成人检查到儿科检查的剂量降低。腹部检查方案中优选使用AEC(成人检查中为84%,儿科检查中为96%)。尽管如此,在成人常规头部检查方案中只有27%使用该系统。数据来自本研究涵盖的诊所和医院,据此估计葡萄牙每1000人每年进行CT检查的频率为206次。葡萄牙人群的估计剂量指数(CTDI)和CT检查频率与其他国家的数据一致。