A Ten-Year Radiostereometric Analysis of Polyethylene Wear Between Oxidized Zirconium and Cobalt Chrome Articulations in Total Hip Arthroplasty.

机构信息

Division of Orthopaedic Surgery, Department of Surgery, London Health Sciences, London, Ontario, Canada.

出版信息

J Arthroplasty. 2022 Jul;37(7S):S692-S696. doi: 10.1016/j.arth.2022.02.099. Epub 2022 Mar 1.

Abstract

BACKGROUND

Oxidized zirconium (OxZi) femoral heads combine the decreased abrasive properties of ceramics with the toughness of metal alloys to lower wear and increase the durability of total hip arthroplasty. The purpose of this study was to compare the 10-year polyethylene wear rates between OxZi/XLPE and cobalt chrome (CoCr)/XLPE articulations.

METHODS

A radiostereometric analysis (RSA) was performed on two cohorts of 23 patients who underwent total hip arthroplasty using either OxZi/XLPE or CoCr/XLPE at a minimum of 10-year follow-up. Cohorts were matched for age, gender, body mass index (BMI), and diagnosis. Polyethylene wear was measured using RSA to determine total and steady-state wear rates for both cohorts. Preoperative and postoperative patient-reported outcome measures (SF12, HHS, and Western Ontario and McMaster Universities Arthritis Index scores) were compared.

RESULTS

The mean total head penetration rate was found to be statistically different between the entire cohorts (OxZi 0.048 ± 0.021 mm/y, CoCr 0.035 ± 0.017 mm/y, P = .02) but not when 28-mm heads only (OxZi 0.045 ± 0.016 mm/y, CoCr 0.034 ± 0.017 mm/y, P = .066) were directly compared. The mean steady-state wear rate was not significantly different between the entire cohorts (OxZi 0.031 ± 0.021 mm/y, CoCr 0.024 ± 0.019 mm/y, P = .24) or 28-mm head cohorts (OxZi 0.028 ± 0.019 mm/y, CoCr 0.024 ± 0.019 mm/y, P = .574). Outcome measures showed no statistical difference except for the Harris Hip Score where the OxZi cohort demonstrated higher median scores.

CONCLUSION

Using RSA to evaluate the 10-year in-vivo head penetration, there was no statistically significant difference in steady-state wear rates between OxZi and CoCr articulations. Both bearing combinations demonstrated wear rates well below the threshold for osteolysis.

摘要

背景

氧化锆(OxZi)股骨头结合了陶瓷的低磨损特性和金属合金的韧性,以降低磨损并提高全髋关节置换术的耐用性。本研究的目的是比较 OxZi/XLPE 和钴铬(CoCr)/XLPE 关节 10 年的聚乙烯磨损率。

方法

对两组 23 例患者进行了放射性立体测量分析(RSA),这些患者在至少 10 年的随访后接受了全髋关节置换术,分别使用 OxZi/XLPE 或 CoCr/XLPE。对两组患者进行了年龄、性别、体重指数(BMI)和诊断匹配。使用 RSA 测量聚乙烯磨损,以确定两组的总磨损率和稳态磨损率。比较术前和术后患者报告的结果测量(SF12、HHS 和西安大略和麦克马斯特大学关节炎指数评分)。

结果

发现整个队列的总头部穿透率存在统计学差异(OxZi 0.048±0.021mm/y,CoCr 0.035±0.017mm/y,P=0.02),但当仅比较 28mm 头时没有统计学差异(OxZi 0.045±0.016mm/y,CoCr 0.034±0.017mm/y,P=0.066)。整个队列的稳态磨损率无显著差异(OxZi 0.031±0.021mm/y,CoCr 0.024±0.019mm/y,P=0.24)或 28mm 头队列(OxZi 0.028±0.019mm/y,CoCr 0.024±0.019mm/y,P=0.574)。除了 OxZi 队列的 Harris 髋关节评分中位数较高外,其他结果测量均无统计学差异。

结论

使用 RSA 评估 10 年体内头部穿透情况,OxZi 和 CoCr 关节在稳态磨损率方面无统计学差异。两种轴承组合的磨损率均远低于溶骨阈值。