Bone Joint J. 2024 Mar 1;106-B(3 Supple A):38-43. doi: 10.1302/0301-620X.106B3.BJJ-2023-0836.R1.
Oxidized zirconium (OxZi) and highly cross-linked polyethylene (HXLPE) were developed to minimize wear and risk of osteolysis in total hip arthroplasty (THA). However, retrieval studies have shown that scratched femoral heads may lead to runaway wear, and few reports of long-term results have been published. The purpose of this investigation is to report minimum ten-year wear rates and clinical outcomes of THA with OxZi femoral heads on HXLPE, and to compare them with a retrospective control group of cobalt chrome (CoCr) or ceramic heads on HXLPE.
From 2003 to 2006, 108 THAs were performed on 96 patients using an OxZi head with a HXLPE liner with minimum ten-year follow-up. Harris Hip Scores (HHS) were collected preoperatively and at the most recent follow-up (mean 13.3 years). Linear and volumetric liner wear was measured on radiographs of 85 hips with a minimum ten-year follow-up (mean 14.5 years). This was compared to a retrospective control group of 45 THAs using ceramic or CoCr heads from October 1999 to February 2005, with a minimum of ten years' follow-up.
Average HHS improved from 50.8 to 91.9 and 51.0 to 89.8 in the OxZi group and control group, respectively (p = 0.644), with no osteolysis in either group. Linear and volumetric wear rates in the OxZi group averaged 0.03 mm/year and 3.46 mm/year, respectively. There was no statistically significant difference in HHS scores, nor in linear or volumetric wear rate between the groups, and no revision for any indication.
The radiological and clinical outcomes, and survivorship of THA with OxZi femoral heads and HXLPE liners, were excellent, and comparable to CoCr or ceramic heads at minimum ten-year follow-up. Wear rates are below what would be expected for development of osteolysis. OxZi-HXLPE is a durable bearing couple with excellent long-term outcomes.
氧化锆(OxZi)和高度交联聚乙烯(HXLPE)的开发旨在最大限度地减少全髋关节置换术(THA)中的磨损和骨溶解风险。然而,检索研究表明,刮擦的股骨头可能导致失控磨损,并且很少有长期结果的报告。本研究旨在报告 OxZi 股骨头在 HXLPE 上的 THA 的最低十年磨损率和临床结果,并将其与钴铬(CoCr)或陶瓷股骨头在 HXLPE 上的回顾性对照组进行比较。
2003 年至 2006 年,96 例患者接受了 108 例 OxZi 股骨头和 HXLPE 衬垫的 THA,随访时间至少为十年。术前和最近随访时(平均 13.3 年)收集 Harris 髋关节评分(HHS)。对 85 例至少随访十年(平均 14.5 年)的髋关节进行 X 线片测量线性和容积衬垫磨损。将其与 1999 年 10 月至 2005 年 2 月期间使用陶瓷或 CoCr 头的 45 例回顾性对照组进行比较,随访时间至少为十年。
OxZi 组和对照组的平均 HHS 分别从 50.8 提高到 91.9 和 51.0 提高到 89.8(p=0.644),两组均无骨溶解。OxZi 组的线性和容积磨损率平均为 0.03mm/年和 3.46mm/年。两组之间的 HHS 评分、线性或容积磨损率均无统计学差异,也没有因任何原因进行翻修。
OxZi 股骨头和 HXLPE 衬垫的 THA 的影像学和临床结果以及生存率均极佳,与 CoCr 或陶瓷头在至少十年的随访中相当。磨损率低于发生骨溶解的预期水平。OxZi-HXLPE 是一种耐用的轴承组合,具有出色的长期结果。