探索性分析不典型圆锥角膜敏感性诊断参数。

An exploratory analysis of forme fruste keratoconus sensitivity diagnostic parameters.

机构信息

School of Biomedical Engineering, Capital Medical University, Beijing, 100069, China.

Department of Medical Engineering, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences, Beijing, 100730, China.

出版信息

Int Ophthalmol. 2022 Aug;42(8):2473-2481. doi: 10.1007/s10792-022-02246-0. Epub 2022 Mar 5.

Abstract

PURPOSE

To secondary statistical analysis of the Pentacam or Corvis ST parameters from literatures, and to obtain more sensitive diagnostic parameters for clinical keratoconus (CKC) and forme fruste keratoconus (FFKC), respectively.

METHODS

The parameters and the corresponding area of ROC curve (AUC) in previous studies were extracted and screened to obtain the database of CKC (Data-CKC) and FFKC (Data-FFKC), respectively. Two different importance evaluation methods (%IncMSE and IncNodePurity) of random forest were used to preliminary select the important parameters. Then, based on the partial dependency analysis, the sensitive diagnostic parameters that had promotion to the diagnostic performance were obtained. Data-FFKC was analyzed in the same way. Finally, a diagnostic test meta-analysis on the sensitive parameter of interest was conducted to verify the reliability of the above analysis methods.

RESULTS

There were 88 parameters with 766 records in Data-CKC, 57 parameters with 346 records in Data-FFKC. Based on two importance evaluation methods, 60 important parameters were obtained, of which 20 were further screened as sensitive parameters of keratoconus, and most of these parameters were related to the thinnest point of cornea. The stiffness parameter at first applanation (SPA1) was the only Corvis ST output parameter sensitive to FFKC except the Tomographic and Biomechanical Index and the Corvis Biomechanical Parameter (CBI). A total of 4 records were included in the meta-analysis of diagnostic tests on SPA1. The results showed that there was threshold effect, but no significant heterogeneity (I = 33%), and the area under the SROC curve was 0.87 (95% CI, 0.84-0.90).

CONCLUSIONS

For the diagnosis of FFKC, the sensitivity of SPA1 is not inferior to the well-known CBI, and may be the earliest Corvis ST output parameter to reflect the changes of corneal biomechanics during keratoconus progression. The elevation parameters based on the typical position of the thinnest point of corneal thickness are of great significance for the diagnosis of keratoconus.

摘要

目的

对 Pentacam 或 Corvis ST 参数进行文献二次统计分析,分别获得更敏感的临床圆锥角膜(CKC)和早期圆锥角膜(FFKC)诊断参数。

方法

提取并筛选先前研究中的参数及其对应的 ROC 曲线面积(AUC),分别获得 CKC 数据库(Data-CKC)和 FFKC 数据库(Data-FFKC)。使用随机森林的两种不同重要性评估方法(%IncMSE 和 IncNodePurity)初步选择重要参数。然后,基于偏依赖分析,获得对诊断性能有促进作用的敏感诊断参数。以同样的方式对 Data-FFKC 进行分析。最后,对感兴趣的敏感参数进行诊断测试荟萃分析,以验证上述分析方法的可靠性。

结果

Data-CKC 中有 88 个参数和 766 个记录,Data-FFKC 中有 57 个参数和 346 个记录。基于两种重要性评估方法,获得了 60 个重要参数,其中 20 个进一步筛选为圆锥角膜的敏感参数,这些参数大多与角膜最薄点有关。首次压平(SPA1)的刚性参数是除 Tomographic and Biomechanical Index 和 Corvis Biomechanical Parameter(CBI)之外,唯一对 FFKC 敏感的 Corvis ST 输出参数。SPA1 的诊断测试荟萃分析共纳入 4 个记录。结果表明存在阈值效应,但无显著异质性(I=33%),SROC 曲线下面积为 0.87(95%CI,0.84-0.90)。

结论

对于 FFKC 的诊断,SPA1 的敏感性并不逊于著名的 CBI,并且可能是最早反映角膜生物力学在圆锥角膜进展过程中变化的 Corvis ST 输出参数。基于角膜最薄点典型位置的隆起参数对圆锥角膜的诊断具有重要意义。