通过数字减法视频密度测定法测量横截面积。

Cross-sectional area measurements by digital subtraction videodensitometry.

作者信息

Simons M A, Kruger R A, Power R L

出版信息

Invest Radiol. 1986 Aug;21(8):637-44. doi: 10.1097/00004424-198608000-00007.

Abstract

A method for measurement of absolute cross-section areas in digital images of iodine-opacified vessels was tested in phantoms containing "normal" segments of circular cross-section and either circular or noncircular "stenoses" over a wide range of iodine concentrations. Accuracy varied with iodine concentration, with a slight tendency to overestimate the area at high concentration (185 mg I/ml) and to underestimate at low concentration (23 mg I/ml). Reproducibility was improved at the higher concentrations, with standard deviation ranging from 2% to 7% of the measured value, compared with 7% to 27% at the lowest concentration. Accuracy and reproducibility of area measurements on either normal or stenotic segments were unchanged by rotation of the vessel by up to 25 degrees out of the plane of the image. This method is easily implemented using computerized image processing techniques and has advantages over either manual or automated edge detection because it eliminates the need for an edge criterion and is independent of the viewing projection.

摘要

一种用于测量碘造影血管数字图像中绝对横截面积的方法,在含有圆形横截面“正常”节段以及各种碘浓度下的圆形或非圆形“狭窄”的模型中进行了测试。准确性随碘浓度而变化,在高浓度(185 mg I/ml)时有轻微高估面积的趋势,而在低浓度(23 mg I/ml)时则有低估的趋势。在较高浓度下可重复性得到改善,标准偏差范围为测量值的2%至7%,而在最低浓度下为7%至27%。血管在图像平面外旋转达25度时,正常或狭窄节段面积测量的准确性和可重复性均未改变。该方法使用计算机图像处理技术易于实现,并且优于手动或自动边缘检测,因为它无需边缘标准且与观察投影无关。