Cardiology & Clinical Research Department, Cardiovascular Institute « Dedinje », Belgrade, Serbia.
Diagnosis and Therapeutic Center, Hôtel Dieu Hospital, Paris, France.
Vasc Health Risk Manag. 2022 Mar 12;18:123-129. doi: 10.2147/VHRM.S353715. eCollection 2022.
Electronic devices for blood pressure (BP) measurements must undergo independent clinical validation as recommended by various authorities and scientific societies.
To assess the accuracy of the Combei BP118A device in the general population according to the Universal Standard Validation Protocol.
The new-developed Combei BP118A device measures BP at the brachial level using the oscillometric method. The study was performed according to the "AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) protocol". Subjects (n = 88) were recruited to fulfill the age, gender, BP and cuff distribution criteria according to the protocol requirements using the same arm sequential BP measurement method. Differences between observers' mercury sphygmomanometer reference measurements and device BP values (test versus reference) and their standard deviation (SD) were calculated.
Ninety-one subjects were selected, 88 of whom were included. The mean BP differences between the simultaneous observers' reference measurements were -0.2 ± 1.7 mmHg for systolic BP (SBP) and -0.2 ± 1.7 mmHg for diastolic BP (DBP). For validation criterion 1, the mean difference ± SD between the reference and device BP values were 3.6 ± 5.5 mmHg for SBP and 1.4 ± 4.5 mmHg for DBP. For criterion 2, the SD of the mean BP differences between the test device and reference BP per subject was 5/4.1 mmHg for SBP/DBP (≤5.9/6.8). These results fulfilled the protocol requirements.
The Combei BP118A oscillometric device for home BP measurement fulfilled all of the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) in general population and consequently can be recommended for home BP measurements.
各种权威机构和科学协会建议,电子血压(BP)测量设备必须进行独立的临床验证。
根据通用标准验证方案评估 Combei BP118A 设备在普通人群中的准确性。
新开发的 Combei BP118A 设备使用振荡测量法在肱动脉水平测量血压。该研究根据“AAMI/ESH/ISO 通用标准(ISO 81060-2:2018)协议”进行。根据协议要求,使用相同手臂顺序血压测量方法,招募受试者(n=88),以满足年龄、性别、血压和袖带分布标准。计算观察者汞柱血压计参考测量值与设备血压值(测试与参考)之间的差异及其标准差(SD)。
选择了 91 名受试者,其中 88 名入选。同时观察者参考测量的平均血压差异为收缩压(SBP)为-0.2±1.7mmHg,舒张压(DBP)为-0.2±1.7mmHg。对于验证标准 1,参考值和设备血压值之间的平均差值±SD 为 SBP 为 3.6±5.5mmHg,DBP 为 1.4±4.5mmHg。对于标准 2,每个受试者测试设备和参考 BP 之间平均 BP 差值的 SD 为 5/4.1mmHg(SBP/DBP)(≤5.9/6.8)。这些结果符合协议要求。
Combei BP118A 家庭血压测量振荡设备符合 AAMI/ESH/ISO 通用标准(ISO 81060-2:2018)的所有要求,因此可以推荐用于家庭血压测量。