THE EFFECT OF OUT-OF-PLANE PATIENT SHIELDING ON CT RADIATION EXPOSURE AND TUBE CURRENT MODULATIONS: A PHANTOM STUDY ACROSS THREE VENDORS.

机构信息

Department of Diagnostic Radiology, Oulu University Hospital, Oulu 90220, Finland.

Department of Radiology, Lapland Central Hospital, Rovaniemi 96101, Finland.

出版信息

Radiat Prot Dosimetry. 2022 Mar 30;198(4):229-237. doi: 10.1093/rpd/ncac032.

Abstract

The aim of this study was to evaluate how out-of-plane patient shielding affects radiation exposure parameters and tube current modulation on different vendors' computed tomography (CT) scanners. Helical CT scans were performed using two homogenous phantoms to mimic patient attenuation. Four CT scanners from three vendors were investigated by varying the distance of the patient shield from the border of the imaging volume. Scans were performed with a shield placed before and after the localizer. Changes in volume computed tomography dose index (CTDIvol), dose-length product (DLP) and tube current-time products were studied. Out-of-field lead shield increased the CTDIvol and DLP values for each scanner at least for one scan setting when the shield was present in the localizer. The most notable changes were recorded with >1.3 pitch values when the shield was closest to the scanned volume (2.5 cm), and the scan direction was towards the shield. The usage of patient shields in the localizer CT scans can disturb TCM even when placed 7.5 cm away from the edge of the scan.

摘要

本研究旨在评估在不同厂商的 CT 扫描仪上,使用体外患者屏蔽对辐射暴露参数和管电流调制的影响。采用两个均匀体模模拟患者衰减,进行螺旋 CT 扫描。通过改变患者屏蔽与成像体积边界的距离,对来自三个厂商的四台 CT 扫描仪进行了研究。在定位器前后放置屏蔽进行扫描。研究了体积 CT 剂量指数(CTDIvol)、剂量长度乘积(DLP)和管电流时间乘积的变化。当定位器中存在屏蔽时,对于每种扫描仪,体外铅屏蔽至少在一种扫描设置中增加了 CTDIvol 和 DLP 值。当屏蔽距离扫描体积最近(2.5cm)且扫描方向朝向屏蔽时,记录到的变化最显著。即使将患者屏蔽放置在距离扫描边缘 7.5cm 处,在定位器 CT 扫描中使用患者屏蔽也可能会干扰 TCM。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验