Kalra Mannudeep K, Dang Pragya, Singh Sarabjeet, Saini Sanjay, Shepard Jo-Anne O
Department of Radiology, Massachusetts General Hospital, Boston, MA 02114, USA.
Korean J Radiol. 2009 Mar-Apr;10(2):156-63. doi: 10.3348/kjr.2009.10.2.156. Epub 2009 Mar 3.
To assess effects of off-centering, automatic exposure control, and padding on attenuation values, noise, and radiation dose when using in-plane bismuth-based shields for CT scanning.
A 30 cm anthropomorphic chest phantom was scanned on a 64-multidetector CT, with the center of the phantom aligned to the gantry isocenter. Scanning was repeated after placing a bismuth breast shield on the anterior surface with no gap and with 1, 2, and 6 cm of padding between the shield and the phantom surface. The "shielded" phantom was also scanned with combined modulation and off-centering of the phantom at 2 cm, 4 cm and 6 cm below the gantry isocenter. CT numbers, noise, and surface radiation dose were measured. The data were analyzed using an analysis of variance.
The in-plane shield was not associated with any significant increment for the surface dose or CT dose index volume, which was achieved by comparing the radiation dose measured by combined modulation technique to the fixed mAs (p > 0.05). Irrespective of the gap or the surface CT numbers, surface noise increased to a larger extent compared to Hounsfield unit (HU) (0-6 cm, 26-55%) and noise (0-6 cm, 30-40%) in the center. With off-centering, in-plane shielding devices are associated with less dose savings, although dose reduction was still higher than in the absence of shielding (0 cm off-center, 90% dose reduction; 2 cm, 61%) (p < 0.0001). Streak artifacts were noted at 0 cm and 1 cm gaps but not at 2 cm and 6 cm gaps of shielding to the surface distances.
In-plane shields are associated with greater image noise, artifactually increased attenuation values, and streak artifacts. However, shields reduce radiation dose regardless of the extent of off-centering. Automatic exposure control did not increase radiation dose when using a shield.
评估在CT扫描中使用平面内铋基屏蔽物时,偏心、自动曝光控制和填充对衰减值、噪声和辐射剂量的影响。
在64排多层螺旋CT上对一个30cm的人体胸部模型进行扫描,模型中心与机架等中心对齐。在前表面放置铋制乳房屏蔽物且屏蔽物与模型表面之间无间隙、有1cm、2cm和6cm的填充物后重复扫描。“屏蔽”后的模型还在机架等中心下方2cm、4cm和6cm处进行模型的组合调制和偏心扫描。测量CT值、噪声和表面辐射剂量。使用方差分析对数据进行分析。
通过将组合调制技术测量的辐射剂量与固定管电流-时间乘积(mAs)进行比较,平面内屏蔽与表面剂量或CT剂量指数容积的任何显著增加均无关(p>0.05)。无论间隙或表面CT值如何,与中心的亨氏单位(HU)(0 - 6cm,26 - 55%)和噪声(0 - 6cm,30 - 40%)相比,表面噪声增加幅度更大。偏心扫描时,平面内屏蔽装置节省的剂量较少,尽管剂量降低仍高于无屏蔽时(偏心0cm,剂量降低90%;2cm,61%)(p<0.0001)。在屏蔽物与表面距离为0cm和1cm间隙时出现条纹伪影,但在2cm和6cm间隙时未出现。
平面内屏蔽与更大的图像噪声、人为增加的衰减值和条纹伪影相关。然而,无论偏心程度如何,屏蔽均可降低辐射剂量。使用屏蔽物时自动曝光控制不会增加辐射剂量。