Department of Radiological Science, College of Health Science, Gachon University, Incheon 21936, Korea.
Department of Health Science, Gachon University Graduate School, Gachon University, Incheon 21936, Korea.
Sensors (Basel). 2022 Mar 16;22(6):2302. doi: 10.3390/s22062302.
This study aimed to develop a magnetic resonance imaging (MRI)-compatible flow delivery system and individualized models of circle of Willis (CoW), which include 50% and 100% blockage in internal carotid artery (ICA50 and ICA100), and 100% blockage in vertebral artery (VA100). Images were obtained using 3D time-of-flight and phase-contrast magnetic resonance angiography (MRA) sequences, and changes in velocity and flow direction at CoW models were analyzed. For the ICA50 and VA100 models, the flow was similar to that of the normal model. For the ICA 50 model, it was found that 50% blockage did not affect cerebral blood flow. For the VA100 model, decreased flow in the posterior cerebral artery and a change to the flow direction in the posterior communicating artery were found. For the ICA100 model, particularly, decreased flow in the ipsilateral middle and anterior cerebral arteries and a change to the flow direction in the ipsilateral anterior cerebral artery of the CoW were found. These results demonstrated that the flow system with various CoW disease models tailored to individual characteristics could be used to predict stroke onset more quickly. For the ICA50 and VA100 models, the possibility of cerebral infarction was significantly lower. On the other hand, for the ICA100 model, there was a high possibility of decreased flow, which could lead to cerebral infarction.
本研究旨在开发一种磁共振成像(MRI)兼容的血流输送系统和个体化的Willis 环(CoW)模型,包括颈内动脉(ICA)50%和 100%阻塞(ICA50 和 ICA100)以及椎动脉(VA)100%阻塞(VA100)。使用 3D 时间飞跃和相位对比磁共振血管造影(MRA)序列获取图像,并分析 CoW 模型中速度和血流方向的变化。对于 ICA50 和 VA100 模型,血流类似于正常模型。对于 ICA50 模型,发现 50%的阻塞不会影响脑血流。对于 VA100 模型,发现大脑后动脉的血流减少以及后交通动脉的血流方向发生变化。对于 ICA100 模型,特别是发现同侧大脑中动脉和大脑前动脉的血流减少以及 CoW 同侧大脑前动脉的血流方向发生变化。这些结果表明,针对个体特征定制的具有各种 CoW 疾病模型的血流系统可用于更快地预测中风发作。对于 ICA50 和 VA100 模型,脑梗死的可能性明显较低。另一方面,对于 ICA100 模型,存在血流减少的高可能性,这可能导致脑梗死。