Correlation between Fundus Autofluorescence and En Face OCT Measurements of Geographic Atrophy.

机构信息

Doheny Image Reading Center, Doheny Eye Institute, Los Angeles, California; Department of Ophthalmology, David Geffen School of Medicine at UCLA, Los Angeles, California.

Doheny Image Reading Center, Doheny Eye Institute, Los Angeles, California; Department of Ophthalmology, David Geffen School of Medicine at UCLA, Los Angeles, California.

出版信息

Ophthalmol Retina. 2022 Aug;6(8):676-683. doi: 10.1016/j.oret.2022.03.017. Epub 2022 Mar 23.

Abstract

PURPOSE

To evaluate the correlation between fundus autofluorescence (FAF) and en face spectral-domain OCT (SD-OCT) measurements of geographic atrophy (GA) area associated with age-related macular degeneration.

DESIGN

Retrospective, cross-sectional study.

PARTICIPANTS

Two hundred seventy eyes from 172 subjects with GA associated with age-related macular degeneration.

METHODS

Subjects with atrophic age-related macular degeneration who underwent both FAF (Heidelberg HRA + Spectralis) and dense volume (128 B-scans over 6 × 6 mm) SD-OCT (Cirrus OCT) imaging were included in this retrospective analysis. The borders of all areas of definite decreased autofluorescence (DDAF) corresponding to GA were manually outlined on FAF images by certified graders at Doheny Image Reading Center using validated planimetric grading tools. Geographic atrophy was also automatically delineated using en face OCT (at the level of the choroid) using an instrument software (Cirrus v.6.2), and segmentation errors were manually corrected before the computation of the GA area. The fundus autofluorescence and SD-OCT-derived measurements were correlated.

MAIN OUTCOME MEASURES

Correlation between the SD-OCT and FAF measurements of the GA area.

RESULTS

The mean GA area measured from the FAF images was 8.1 ± 5.04 mm, compared with an automated, uncorrected GA area of 6.82 ± 3.84 mm measured using SD-OCT. Despite the presence of apparent OCT segmentation errors, there was a significant correlation between the FAF and uncorrected SD-OCT measurements (r = 0.80; P < 0.001). After the manual correction of the SD-OCT GA segmentation errors, the measured GA area increased to 7.29 ± 4.18 mm, and the correlation with the FAF-determined GA area significantly improved (r = 0.98; P < 0.001).

CONCLUSIONS

Spectral-domain OCT-derived measurements of GA correlate well with areas of DDAF obtained from FAF images. The manual correction of SD-OCT segmentation errors can further improve this correlation. These observations may support the use of SD-OCT-based measurements of the GA area in clinical research and clinical trials.

摘要

目的

评估与年龄相关性黄斑变性相关的地图状萎缩(GA)区域的眼底自发荧光(FAF)和共焦光谱域 OCT(SD-OCT)测量值之间的相关性。

设计

回顾性、横断面研究。

参与者

270 只眼睛来自 172 名与年龄相关性黄斑变性相关的 GA 患者。

方法

纳入了在海德堡 HRA+Spectralis 进行 FAF(眼底自发荧光)和密集体积(6×6mm 上的 128 个 B 扫描)SD-OCT(Cirrus OCT)成像的具有萎缩性年龄相关性黄斑变性的患者。由 Doheny Image Reading Center 的认证分级员使用经过验证的平面分级工具在 FAF 图像上手动描绘出与 GA 相对应的所有明确的自发荧光减少区域(DDAF)的边界。使用仪器软件(Cirrus v.6.2)在脉络膜水平自动描绘 GA 的 OCT 图像,并在计算 GA 面积之前手动纠正分段错误。对 FAF 和 SD-OCT 测量值进行相关性分析。

主要观察指标

SD-OCT 和 FAF 测量 GA 面积之间的相关性。

结果

FAF 图像测量的 GA 平均面积为 8.1±5.04mm,而使用 SD-OCT 自动测量的未校正 GA 面积为 6.82±3.84mm。尽管存在明显的 OCT 分段错误,但 FAF 和未校正的 SD-OCT 测量值之间存在显著相关性(r=0.80;P<0.001)。在手动校正 SD-OCT GA 分段错误后,测量的 GA 面积增加到 7.29±4.18mm,与 FAF 确定的 GA 面积的相关性显著改善(r=0.98;P<0.001)。

结论

SD-OCT 测量的 GA 与 FAF 图像中获得的 DDAF 区域相关性良好。SD-OCT 分段错误的手动校正可以进一步提高这种相关性。这些观察结果可能支持在临床研究和临床试验中使用基于 SD-OCT 的 GA 面积测量值。