Does cerebral oximetry always measure brain tissue oxygen saturation? An anatomical study utilizing computed tomography.

作者信息

Naftalovich Rotem, Chyu Darrick, Denny John T, Hasan Aysha, Pantin Enrique J

机构信息

Medical Corps of the U.S. Army, U.S. Army Medical Department, 3630 Stanley Road, STE 104, Fort Sam Houston, TX, USA 78234.

Department of Anesthesiology and Perioperative Care, Rutgers - New Jersey Medical School, Medical Science Building, 185 South Orange Avenue Newark, NJ, USA 07103.

出版信息

J Anaesthesiol Clin Pharmacol. 2021 Oct-Dec;37(4):537-541. doi: 10.4103/joacp.JOACP_395_19. Epub 2022 Jan 6.

Abstract

BACKGROUND AND AIMS

to quantify the scalp-cortex distance and determine its variation among patients. We hypothesized that in a significant number of patients, this distance is greater than the maximum penetration depth of current cerebral oximeters.

MATERIAL AND METHODS

A retrospective anatomic study using transverse head CT images selected randomly from 102 patients over the age of 18 years without brain swelling, intracranial mass effect, or brain hemorrhage. Scalp-cortex distances were determined at two separate locations along the craniocaudal axis; most cephalad to the frontal sinus (I) and also 2 cm cephalad to that location (I). Multiple measurements were obtained bilaterally at 1, 3, 5, 7, and 9 cm from midline.

RESULTS

The average scalp-cortex distance was 14.3 mm and 15 mm at I and I respectively. Distances varied more in I than in I; from the measurements, 12.8% . 6.8% were over 20 mm, 4.4% . 2.2% over 25 mm, 1.1% . 0.6% over 35 mm and 0.6% . none over 40 mm at I and I respectively. 1.5% of the measurements at I were over 30 mm.

CONCLUSION

Cerebral oximetry manufacturers all claim to measure cerebral tissue up to a depth of 20-25 mm; 20 mm with the EQUANOX and INVOS compared with 25 mm with the FORE-SIGHT. Scalp-cortex distance is within 25 mm in more than 95% of patients. However, even with the probe placed as per the manufacturer's recommendations, in a small but significant subset of patients, this distance is greater than the maximum penetration depth of current cerebral oximeters and hence may not reflect actual brain tissue oxygen saturation.

摘要

背景与目的

量化头皮与皮层之间的距离,并确定患者之间该距离的变化情况。我们假设,在相当数量的患者中,这一距离大于当前脑血氧饱和度仪的最大穿透深度。

材料与方法

一项回顾性解剖学研究,使用从102名18岁以上无脑肿胀、颅内占位效应或脑出血的患者中随机选取的头颅横断CT图像。在沿头尾轴的两个不同位置测定头皮与皮层之间的距离;最靠近额窦的位置(I)以及该位置头侧2厘米处(I)。在距中线1、3、5、7和9厘米处双侧进行多次测量。

结果

I处和I处的平均头皮与皮层距离分别为14.3毫米和15毫米。I处的距离变化比I处更大;从测量结果来看,I处和I处分别有12.8%、6.8%超过20毫米,4.4%、2.2%超过25毫米,1.1%、0.6%超过35毫米,0.6%、无超过40毫米。I处1.5%的测量值超过30毫米。

结论

脑血氧饱和度仪制造商均声称能够测量深度达20 - 25毫米的脑组织;EQUANOX和INVOS为20毫米,而FORE - SIGHT为25毫米。超过95%的患者头皮与皮层距离在25毫米以内。然而,即使按照制造商的建议放置探头,在一小部分但数量可观的患者中,这一距离大于当前脑血氧饱和度仪的最大穿透深度,因此可能无法反映实际的脑组织氧饱和度。