Naftalovich Rotem, Chyu Darrick, Denny John T, Hasan Aysha, Pantin Enrique J
Medical Corps of the U.S. Army, U.S. Army Medical Department, 3630 Stanley Road, STE 104, Fort Sam Houston, TX, USA 78234.
Department of Anesthesiology and Perioperative Care, Rutgers - New Jersey Medical School, Medical Science Building, 185 South Orange Avenue Newark, NJ, USA 07103.
J Anaesthesiol Clin Pharmacol. 2021 Oct-Dec;37(4):537-541. doi: 10.4103/joacp.JOACP_395_19. Epub 2022 Jan 6.
to quantify the scalp-cortex distance and determine its variation among patients. We hypothesized that in a significant number of patients, this distance is greater than the maximum penetration depth of current cerebral oximeters.
A retrospective anatomic study using transverse head CT images selected randomly from 102 patients over the age of 18 years without brain swelling, intracranial mass effect, or brain hemorrhage. Scalp-cortex distances were determined at two separate locations along the craniocaudal axis; most cephalad to the frontal sinus (I) and also 2 cm cephalad to that location (I). Multiple measurements were obtained bilaterally at 1, 3, 5, 7, and 9 cm from midline.
The average scalp-cortex distance was 14.3 mm and 15 mm at I and I respectively. Distances varied more in I than in I; from the measurements, 12.8% . 6.8% were over 20 mm, 4.4% . 2.2% over 25 mm, 1.1% . 0.6% over 35 mm and 0.6% . none over 40 mm at I and I respectively. 1.5% of the measurements at I were over 30 mm.
Cerebral oximetry manufacturers all claim to measure cerebral tissue up to a depth of 20-25 mm; 20 mm with the EQUANOX and INVOS compared with 25 mm with the FORE-SIGHT. Scalp-cortex distance is within 25 mm in more than 95% of patients. However, even with the probe placed as per the manufacturer's recommendations, in a small but significant subset of patients, this distance is greater than the maximum penetration depth of current cerebral oximeters and hence may not reflect actual brain tissue oxygen saturation.
量化头皮与皮层之间的距离,并确定患者之间该距离的变化情况。我们假设,在相当数量的患者中,这一距离大于当前脑血氧饱和度仪的最大穿透深度。
一项回顾性解剖学研究,使用从102名18岁以上无脑肿胀、颅内占位效应或脑出血的患者中随机选取的头颅横断CT图像。在沿头尾轴的两个不同位置测定头皮与皮层之间的距离;最靠近额窦的位置(I)以及该位置头侧2厘米处(I)。在距中线1、3、5、7和9厘米处双侧进行多次测量。
I处和I处的平均头皮与皮层距离分别为14.3毫米和15毫米。I处的距离变化比I处更大;从测量结果来看,I处和I处分别有12.8%、6.8%超过20毫米,4.4%、2.2%超过25毫米,1.1%、0.6%超过35毫米,0.6%、无超过40毫米。I处1.5%的测量值超过30毫米。
脑血氧饱和度仪制造商均声称能够测量深度达20 - 25毫米的脑组织;EQUANOX和INVOS为20毫米,而FORE - SIGHT为25毫米。超过95%的患者头皮与皮层距离在25毫米以内。然而,即使按照制造商的建议放置探头,在一小部分但数量可观的患者中,这一距离大于当前脑血氧饱和度仪的最大穿透深度,因此可能无法反映实际的脑组织氧饱和度。