To investigate the 5-year clinic outcomes of rigid iris-fixated pIOL on the visual performance, complications and intraocular light scattering in northern Chinese.
Thirty eyes implanted with iris-fixated phakic IOLs (pIOLs group) and 34 eyes with high myopia (myopia group) were involved in this study. At preoperatively, 6 months, 1, 3, and 5 years postoperatively, the uncorrected visual acuity (UCVA), best-spectacle corrected visual acuity (BSCVA), endothelial cell density (ECD), and intraocular pressure (IOP) were measured in pIOLs group. The objective scatter index (OSI), modulation transfer function cut off (MTF cut off), and Strehl ratio were measured by Optical Quality Analysis System and a pseudophakic dysphotopsia questionnaire (PDQ) was used to evaluate the subjects' satisfaction in pIOLs and myopia groups.
At 5 years postoperatively, an UCVA of 20/20 or better was found in 43.33% of eyes in pIOLs group. At 6 months, 1, 3, and 5 years postoperatively, the mean ECD decrease were 1.29% ± 0.45%, 2.59% ± 1.30%, 6.67% ± 2.26%, and 10.80% ± 3.48%. The value of OSI in pIOLs group was significantly higher than that in myopia group (P < 0.001). The PDQ results showed that the subjects in myopia group complained less with intolerance of bright lights than those in pIOLs group. The values of Strehl ratio and MTF cut off in pIOLs group were significantly lower than that in myopia group (P < 0.001).
Iris-fixated pIOL induce more intraocular light scattering. A significant decrease in ECD was observed at 5 years postoperatively. An annual evaluation of ECD is necessary for patients undergoing pIOL implantation.