Measuring Foot Abduction Brace Wear Time Using a Single 3-Axis Accelerometer.

机构信息

School of Health and Society, University of Salford, Salford M5 4WT, UK.

Shaare Zedek Medical Center, Jerusalem 9103102, Israel.

出版信息

Sensors (Basel). 2022 Mar 22;22(7):2433. doi: 10.3390/s22072433.

Abstract

The recommended treatment for idiopathic congenital clubfoot deformity involves a series of weekly castings, surgery, and a period of bracing using a foot abduction brace (FAB). Depending on the age of the child, the orthotic should be worn for periods that reduce in duration as the child develops. Compliance is vital to achieve optimal functional outcomes and reduce the likelihood of reoccurrence, deformity, or the need for future surgery. However, compliance is typically monitored by self-reporting, which is time-consuming to implement and lacks accuracy. This study presents a novel method for objectively monitoring FAB wear using a single 3-axis accelerometer. Eleven families mounted an accelerometer on their infant's FAB for up to seven days. Parents were also given a physical diary that was used to record the daily application and removal of the orthotic in line with their treatment. Both methods produced very similar measurements of wear that visually aligned with the movement measured by the accelerometer. Bland Altman plots showed a -0.55-h bias in the diary measurements and the limits of agreement ranging from -2.96 h to 1.96 h. Furthermore, the Cohens Kappa coefficient for the entire dataset was 0.88, showing a very high level of agreement. The method provides an advantage over existing objective monitoring solutions as it can be easily applied to existing FABs, preventing the need for bespoke monitoring devices. The novel method can facilitate increased research into FAB compliance and help enable FAB monitoring in clinical practice.

摘要

特发性先天性马蹄内翻足畸形的推荐治疗方法包括一系列每周一次的石膏固定、手术以及使用足部外展支具(foot abduction brace,FAB)进行一段时间的支具固定。根据儿童的年龄,支具的佩戴时间会逐渐缩短,因为儿童在发育。为了达到最佳的功能效果并降低复发、畸形或未来需要手术的可能性,依从性至关重要。然而,依从性通常通过自我报告来监测,这种方法既耗时又缺乏准确性。本研究提出了一种使用单个三轴加速度计客观监测 FAB 佩戴的新方法。11 个家庭将加速度计安装在婴儿的 FAB 上,最长佩戴了七天。父母还收到了一份物理日记,用于记录矫形器的日常应用和去除情况,与他们的治疗方案保持一致。这两种方法都非常相似地测量了佩戴情况,与加速度计测量的运动情况相符。Bland-Altman 图显示,日记测量值存在 -0.55 小时的偏差,一致性界限范围为 -2.96 小时至 1.96 小时。此外,整个数据集的 Cohen's Kappa 系数为 0.88,显示出非常高的一致性。该方法与现有的客观监测解决方案相比具有优势,因为它可以轻松应用于现有的 FAB,无需定制监测设备。该新方法可以促进对 FAB 依从性的更多研究,并有助于在临床实践中实现 FAB 监测。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/2ffc/9003136/687d38a5522e/sensors-22-02433-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验