CT body stereotaxic system for placement of needle arrays.

作者信息

Onik G, Cosman E, Wells T, Goldberg H I, Moss A, Kane R, Moore S, Stauffer P, Costello P, Hoddick W

出版信息

Int J Radiat Oncol Biol Phys. 1987 Jan;13(1):121-8. doi: 10.1016/0360-3016(87)90269-0.

Abstract

A new CT body stereotaxic system that is particularly suited to multiple placements of needles precisely in parallel is described. By scanning through a right triangle placed on the patient's skin, the stereotaxic method defines an entry point for the first needle placement. An articulating arm is then used to aim the needle at the entry point and hold the needle at the correct angle. The arm can be angled so that the complex approaches from one scan plane to another can be made to place needle arrays through inaccessible lesions, for instance, beneath the diaphragm. Animal and phantom studies have shown that placement of multiple needle arrays in parallel from CT scan data is possible.