Suppr超能文献

氧化锆股骨部件骨折在全膝关节置换术后。

Fracture of the oxidized zirconium femoral component after total knee arthroplasty.

机构信息

Minamikawa Orthopaedic Hospital, Fukuoka, Japan.

Minamikawa Orthopaedic Hospital, Fukuoka, Japan.

出版信息

Knee. 2022 Jun;36:27-32. doi: 10.1016/j.knee.2022.03.014. Epub 2022 Apr 18.

Abstract

BACKGROUND

Femoral component fracture is a rare complication of total knee arthroplasty (TKA).

CASE

We report a case of oxidized zirconium (Oxinium) femoral component fracture after total knee arthroplasty. The fracture site was the junction of the central and medial flanges. The patellar component and polyethylene insert had delamination at the contact point of the fracture line, and the tibial tray had loosening at the medial side. There was no cement adherence at the component fracture site, suggesting that debonding had occurred at the cement-implant interface in this area. Examination with a scanning electron microscope revealed beach marks, which are characteristic findings of metal fatigue.

CONCLUSION

We considered that the cause of femoral component fracture was a fatigue fracture due to poor fixation of the component to the bone caused by poor osteotomy technique or poor cementing technique. To our knowledge, this is the first case of Oxinium femoral component fracture.

摘要

背景

股骨部件骨折是全膝关节置换术(TKA)的一种罕见并发症。

病例报告

我们报告了一例全膝关节置换术后氧化锆(Oxinium)股骨部件骨折。骨折部位位于中央和内侧法兰的交界处。髌骨部件和聚乙烯插入物在骨折线的接触点处分层,胫骨托内侧松动。在部件骨折部位没有水泥附着,表明该区域的水泥-植入物界面发生了脱粘。扫描电子显微镜检查显示沙滩标记,这是金属疲劳的特征性发现。

结论

我们认为股骨部件骨折的原因是由于部件与骨骼的固定不良,导致部件固定不良或水泥技术不佳,导致疲劳骨折。据我们所知,这是首例 Oxinium 股骨部件骨折。